To evaluate the reproducibility of the imo binocular random single-eye test (BRSET) and Humphrey Field Analyzer (HFA) monocular test in patients with glaucoma.
Retrospective observational study.
We measured the visual fields (VF) of patients with glaucoma using the BRSET and HFA. All tests were repeated two months later. Mean sensitivity (MS), mean deviation (MD), sensitivity at each test location, and reliability indices were compared between the test days. Wilcoxon signed-rank test, interclass correlation coefficient (ICC), correlation coefficients, and Bland-Altman plots were generated for analysis.
We analyzed the VFs of 46 patients with glaucoma. There were no test-retest differences for MS and MD, and ICCs were > 0.9 for MS and MD in both perimeters. Inter-test correlations for MS and MD were high. The limits of agreement (LoAs) (lower, upper limit) between test days for MS were (- 3.4, 4.0) for BRSET and (-3.3, 3.0) for HFA. The LoA for MD was (- 3.3, 3.8) for BRSET and (- 3.2, 2.9) for HFA. Sensitivity at each testing location was more variable between testing days for BRSET than for HFA. For reliability indices, LoAs between testing days were wider for BRSET than for HFA.
The imo BRSET showed similar reproducibility to HFA in MS and MD. However, sensitivity at each test location varied more for BRSET than for HFA. Further studies are needed to verify the reproducibility of the imo BRSET.
评估 imo 双目随机单眼测试(BRSET)和 Humphrey 视野分析仪(HFA)单眼测试在青光眼患者中的可重复性。
回顾性观察性研究。
我们使用 BRSET 和 HFA 测量青光眼患者的视野(VF)。所有测试均在两个月后重复。比较两次测试日之间的平均敏感度(MS)、平均偏差(MD)、各测试点的敏感度和可靠性指数。采用 Wilcoxon 符号秩检验、组内相关系数(ICC)、相关系数和 Bland-Altman 图进行分析。
我们分析了 46 例青光眼患者的 VF。MS 和 MD 无测试-再测试差异,BRSET 和 HFA 中 MS 和 MD 的 ICC 均>0.9。MS 和 MD 的测试间相关性较高。MS 的测试日之间的界限(下限、上限)为 BRSET(-3.4, 4.0)和 HFA(-3.3, 3.0)。MD 的 BRSET 为(-3.3, 3.8),HFA 为(-3.2, 2.9)。BRSET 各测试点的敏感度在两次测试日之间的变异性大于 HFA。对于可靠性指数,BRSET 的测试日之间的界限(下限、上限)比 HFA 更宽。
imo BRSET 在 MS 和 MD 方面与 HFA 具有相似的可重复性。然而,BRSET 各测试点的敏感度在两次测试日之间的变化大于 HFA。需要进一步的研究来验证 imo BRSET 的可重复性。