Evaluation of average glandular dose (AGD) in screening and diagnostic digital mammography and digital breast tomosynthesis (DBT) towards establishing a reference dose range band (DRB): a developing country experience.

机构信息

Department of Nuclear Science, University of Colombo, Colombo 00300, Sri Lanka.

Department of Radiology, National Hospital of Sri Lanka, Colombo 01000, Sri Lanka.

出版信息

J Radiol Prot. 2023 Jul 26;43(3). doi: 10.1088/1361-6498/ace833.

Abstract

This study represents the first national survey conducted in Sri Lanka to establish national diagnostic reference levels (NDRLs) for screening and diagnostic acquisitions in digital mammography (2D-DM) and digital breast tomosynthesis (DBT). Additionally, the study investigated the relationship between average glandular dose (AGD) and compressed breast thickness (CBT) and introduced a novel concept called dose range bands (DRBs) as a tool for optimizing radiation dose in mammography. DICOM metadata was used to extract dose data and exposure parameters from women undergoing either screening (350) or diagnostic (750) DM. The analysis included both craniocaudal and mediolateral oblique views of each breast, acquired using 2D-DM and DBT imaging techniques. The NDRL (AGD per view) was 1.97 mGy and 2.01 mGy for diagnostic DM and DBT, respectively. The corresponding NDRLs for screening acquisition were 2.44 mGy and 2.30 mGy. The mean DBT/DM AGD ratio was 1.39 and 0.97 for diagnostic and screening, respectively. Further, the association between the average glandular AGD per view and CBT is stronger in DBT than in 2D-DM. The study findings highlight the need for standardisation of compression practices, considering factors such as the radiographer's experience, imaging equipment, breast density, age, breast size, and pain threshold.

摘要

本研究代表了斯里兰卡首次进行的全国性调查,旨在为数字乳腺摄影(2D-DM)和数字乳腺断层合成(DBT)的筛查和诊断采集建立国家诊断参考水平(NDRL)。此外,该研究还调查了平均腺体剂量(AGD)和压缩乳腺厚度(CBT)之间的关系,并引入了一个新的概念,称为剂量范围带(DRB),作为优化乳腺摄影中辐射剂量的工具。DICOM 元数据用于从接受筛查(350 例)或诊断(750 例)DM 的女性中提取剂量数据和曝光参数。分析包括使用 2D-DM 和 DBT 成像技术采集的每侧乳腺的头尾位和内外斜位。NDRL(每视图的 AGD)分别为诊断 DM 和 DBT 的 1.97 mGy 和 2.01 mGy。筛查采集的相应 NDRL 分别为 2.44 mGy 和 2.30 mGy。诊断和筛查的 DBT/DM AGD 比值平均值分别为 1.39 和 0.97。此外,与 CBT 相比,DBT 中每视图的平均腺体 AGD 与 CBT 之间的关联更强。研究结果强调了标准化压缩实践的必要性,需要考虑放射技师的经验、成像设备、乳腺密度、年龄、乳房大小和疼痛阈值等因素。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验