Block F E, Detko G J
J Clin Monit. 1986 Jul;2(3):203-5. doi: 10.1007/BF01620553.
Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the "alarm off" function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.
由于Nellcor N - 100脉搏血氧仪通常对电灼干扰相当敏感,可能会显示错误的血氧饱和度值,并且可能会发出可听见的误报声。我们通过反复试验发现了几种可将干扰和误报都降至最低的技术。将脉搏血氧仪传感器和血氧仪主机尽可能远离手术部位和电灼接地极板,通常会减少干扰。如果这一步还不够,可通过将高脉搏率警报设置为约160、在模式2(快速响应模式)下操作设备以及将“警报关闭”功能设置为120秒来降低误报发生率。这些步骤通常会显著降低误报发生率。