Suppr超能文献

医疗机构中输液设备的流速准确性:一项系统综述

Flow rate accuracy of infusion devices within healthcare settings: a systematic review.

作者信息

Atanda Opeyemi, West Jonathan, Stables Tom, Johnson Chris, Merrifield Robert, Kinross James

机构信息

Department of Surgery and Cancer, Faculty of Medicine, Imperial College London, St. Mary Campus, QEQM Building, London, SW7 2AZ, UK.

Flomark Limited, UK.

出版信息

Ther Adv Drug Saf. 2023 Jul 21;14:20420986231188602. doi: 10.1177/20420986231188602. eCollection 2023.

Abstract

BACKGROUND

One in five patients admitted to the hospital treated with intravenous (IV) fluid therapy suffer complications due to inappropriate administration. Errors have been reported in 13-84% of the preparation and administration of IV medications. The safe delivery of IV fluids requires precise rate administration.

OBJECTIVES

This systematic review aims to determine the accuracy of infusion sets and devices and examine the factors that affect the flow rate accuracy of devices.

DATA SOURCES AND METHODS

Six databases (CINAHL, MEDLINE PubMed, EMBASE, Web of Science and Cochrane Database of systematic reviews) were systematically searched. Search terms included infusion pumps, infusion devices, flow rate accuracy, fluid administration rate, gravity-led infusion set and fluid balance. Studies were included if they examined infusion devices' flow rate accuracy and drop rates for fluids or non-oncological drugs. Findings were tabulated and synthesised qualitatively. The quality of the studies was examined based on the design of the studies due to their heterogeneity.

RESULTS

Eight studies were included: Four studies were conducted on human subjects in the hospital environment; studies recruited 182 participants between the ages of 18 and 94 years. Two studies examined flow rate accuracy in recruited patients across 509 observations and 2387 drip hours. No trials prospectively assessed the accuracy of infusion pumps in the clinical domain, and no studies were reported on patient safety outcomes. Four studies examined the impact of mechanical and physiological factors on the flow rate accuracies of infusion devices. Height and back pressure simulated vibrating conditions, the viscosity of IV fluid and the positions of patients were reported to have a significant impact on infusion volume and flow rates of infusion devices. Additionally, giving sets that vary from the manufacturer's specifications are reported to increase error percent by 10-20%.

CONCLUSION

Infusion devices are an important source of error in administering IV fluids. Yet, there needs to be more prospective trial data to support their clinical accuracy and the impact on patient outcomes. Future flow variability and accuracy studies should capture their impact on patient safety and clinical outcomes.

摘要

背景

在接受静脉输液治疗的住院患者中,五分之一的患者因给药不当而出现并发症。据报道,静脉用药的配制和给药过程中出现错误的比例在13%至84%之间。安全输注静脉输液需要精确控制输注速度。

目的

本系统评价旨在确定输液器和装置的准确性,并研究影响装置流速准确性的因素。

数据来源与方法

系统检索了六个数据库(CINAHL、MEDLINE PubMed、EMBASE、科学网和Cochrane系统评价数据库)。检索词包括输液泵、输液装置、流速准确性、液体给药速度、重力输液器和液体平衡。如果研究考察了输液装置的流速准确性以及液体或非肿瘤药物的滴速,则纳入研究。研究结果进行列表并定性综合。由于研究的异质性,根据研究设计对研究质量进行了评估。

结果

纳入八项研究:四项研究在医院环境中对人类受试者进行;研究招募了182名年龄在18至94岁之间的参与者。两项研究在509次观察和2387滴注小时内考察了招募患者的流速准确性。没有试验前瞻性评估临床领域输液泵的准确性,也没有关于患者安全结果的研究报告。四项研究考察了机械和生理因素对输液装置流速准确性的影响。据报道,高度和背压模拟振动条件、静脉输液的粘度以及患者的体位对输液装置的输液量和流速有显著影响。此外,据报道,与制造商规格不同的输液器会使误差百分比增加10%至20%。

结论

输液装置是静脉输液给药错误的一个重要来源。然而,需要更多的前瞻性试验数据来支持其临床准确性以及对患者结局的影响。未来关于流速变异性和准确性的研究应捕捉其对患者安全和临床结局的影响。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/67d9/10363896/37b0fd5210f0/10.1177_20420986231188602-fig1.jpg