Fidler M W
Spine (Phila Pa 1976). 1986 May;11(4):362-6. doi: 10.1097/00007632-198605000-00013.
Postmortem examination of a spine, which had been instrumented anteriorly at one level and posteriorly at another following vertebral body resections, showed that the posterior instrumentation was more effective in resisting torsion. Postmortem examination of a second case, where the anterior fixation had been strengthened by the addition of cement studs in the vertebral end-plates and a paravertebral modified and strengthened Zielke screw/rod system, showed this second anterior system to be an improvement for resisting torsion following the resection of one vertebral body. Anterior interbody devices become enveloped in a firm smooth fibrous capsule. A preliminary experiment on a cadaver spine had shown that anterior interbody fixation was better than posterior instrumentation for resisting flexion-compression forces.