Wedmore, Somerset, United Kingdom.
East Lydford, Somerset, United Kingdom.
Br Dent J. 2023 Jul;235(2):105-111. doi: 10.1038/s41415-023-6061-9.
It is essential to ensure that reusable dental instruments are scrupulously clean as a first step in the decontamination process. Any residual soil on the surface of equipment creates a risk that will prevent steam, generated during sterilisation, from condensing on the surface of the instrument and raising the temperature to that required to ensure sterilisation. Further, most microorganisms will struggle to attach to the surface of a clean instrument. Adherence of microorganisms to the surface of an instrument that is contaminated with organic soil or the irregular surface of residual dental materials is likely to be greatly enhanced. It is therefore important to ensure that residual dental materials are removed from the surface of instruments, at the point of use, before they are allowed to set.It is also important to ensure that the methods used to clean dental equipment do not damage the instrument. Any method that has the potential to scratch the surface of an instrument and therefore make it more difficult to clean should be avoided. Aggressive cleaning can lead to removal of the passivation layer, the chromium-rich layer that prevents corrosion occurring, formed on the surface of instruments made from stainless steel. If corrosion forms on the surface of an instrument, the surface becomes roughened and possibly stained. Surface staining can then be difficult to differentiate from other stains or marks caused by retained biological soil. Corrosion on the surface of an instrument will make the effective cleaning of the surface more difficult and therefore corroded or stained instruments should be discarded.
确保可重复使用的牙科器械彻底清洁是消毒过程的第一步。设备表面的任何残留污垢都会产生风险,从而阻止蒸汽在消毒过程中凝结在器械表面并将温度提高到确保消毒所需的温度。此外,大多数微生物很难附着在清洁器械的表面上。如果器械表面受到有机污垢或残留牙料不规则表面的污染,微生物的附着很可能大大增强。因此,在允许残留牙料在器械表面凝固之前,务必在使用点从器械表面清除残留牙料。确保用于清洁牙科设备的方法不会损坏器械也很重要。任何可能划伤器械表面从而更难以清洁的方法都应避免。剧烈清洁会导致不锈钢制成的器械表面形成的富铬钝化层(防止腐蚀的形成)被去除。如果器械表面发生腐蚀,表面会变得粗糙并可能变色。然后,表面染色可能难以与其他因残留生物污垢引起的染色或标记区分开来。器械表面的腐蚀会使表面的有效清洁更加困难,因此应丢弃腐蚀或染色的器械。