反复灭菌和超声清洗对刮治器刀片的影响。

Effect of repeated sterilization and ultrasonic cleaning on curet blades.

作者信息

Stach D J, Cross-Poline G N, Newman S M, Tilliss T S

机构信息

Department of Dental Hygiene, University of Colorado School of Dentistry, Denver, Colorado, USA.

出版信息

J Dent Hyg. 1995 Jan-Feb;69(1):31-9.

PMID:8632200
Abstract

PURPOSE

Instrument cleaning and sterilization are important steps in infection control. This study investigated the effect of repeated ultrasonic cleaning, cycles of autoclave or chemiclave, on the surface and cutting edge of stainless steel and carbon steel curets.

METHODS

Eight carbon steel and six stainless steel Columbia #13/14 curets were cleaned or sterilized in a series of treatments by one of the following methods: ultrasonic cleaning, chemiclave sterilization, and autoclave sterilization. In addition, two carbon steel instruments were placed in an anticorrosive dip followed by autoclave. The blades of the curets were examined and photographed with the scanning electron microscope (SEM) at 200 X and 1000 X for baseline and after 2, 4, 8, 16, and 32 treatment cycles. Photographs were evaluated for visible change in surface appearance by four examiners independently. Photographs for each treatment time were compared to the baseline (pretreatment) photographs for the same instrument. They were scored as (1) no visible change; (2) slight change; or (3) moderate to major change. Changes observed were: surface pitting, corrosion products as additions to the surface, edge deterioration, or loss of structure.

RESULTS

Stainless steel curets showed slight or no change after the three treatments. Carbon steel curets were affected by all treatments. The chemiclave produced slight change. The autoclave produced slight change after the fourth treatment cycle and moderate to major change after the eighth cycle. Those carbon steel instruments treated by anticorrosive dip followed by the autoclave showed pronounced change after 16 or 32 cycles. The most varied results were from the ultrasonic cleaner; one blade showed moderate to major change after only four cycles, the other showed slight change at that point and major change only after 32 cycles.

CONCLUSION

With stainless steel curets, ultrasonic cleanings or sterilization with autoclave or chemiclave can be used without visibly affecting the cutting edge. With carbon steel instruments, chemiclave is the least damaging sterilization method followed by anticorrosive treatment before autoclaving. Use of the autoclave without the anticorrosive pretreatment or use of an ultrasonic cleaner negatively affects the integrity of the surface and cutting edge of these instruments.

摘要

目的

器械清洁与消毒是感染控制中的重要环节。本研究调查了重复超声清洗、高压蒸汽灭菌或化学灭菌循环对不锈钢和碳钢刮匙表面及刃口的影响。

方法

八把碳钢和六把不锈钢的哥伦比亚13/14号刮匙通过以下方法之一进行一系列的清洁或消毒处理:超声清洗、化学灭菌、高压蒸汽灭菌。此外,两把碳钢器械先进行防腐浸泡,然后再进行高压蒸汽灭菌。在基线状态以及2、4、8、16和32个处理循环后,用扫描电子显微镜(SEM)在200倍和1000倍下对刮匙的刃口进行检查和拍照。由四名检查人员独立评估照片中表面外观的可见变化。将每个处理时间的照片与同一器械的基线(预处理)照片进行比较。评分标准为:(1)无可见变化;(2)轻微变化;或(3)中度至重度变化。观察到的变化包括:表面点蚀、表面附着腐蚀产物、刃口劣化或结构丧失。

结果

三种处理后,不锈钢刮匙显示轻微变化或无变化。碳钢刮匙受到所有处理的影响。化学灭菌产生轻微变化。高压蒸汽灭菌在第四个处理循环后产生轻微变化,在第八个循环后产生中度至重度变化。那些先进行防腐浸泡再进行高压蒸汽灭菌的碳钢器械在16或32个循环后显示出明显变化。超声清洗器的结果差异最大;一把刃口仅在四个循环后就显示出中度至重度变化,另一把在此时显示轻微变化且仅在32个循环后才显示出重度变化。

结论

对于不锈钢刮匙,超声清洗或高压蒸汽灭菌或化学灭菌不会明显影响刃口。对于碳钢器械,化学灭菌是损害最小的灭菌方法,其次是在高压蒸汽灭菌前进行防腐处理。未经防腐预处理使用高压蒸汽灭菌器或使用超声清洗器会对这些器械的表面完整性和刃口产生负面影响。