Suppr超能文献

[使用灯芯导管测量筋膜间隔压力]

[Compartment pressure measurements with a wick catheter].

作者信息

Festge O A, Gross W, Tischer W

出版信息

Zentralbl Chir. 1986;111(11):674-83.

PMID:3751367
Abstract

There is no generally accepted method for compartmental pressure measurement, as yet. Substantial disadvantages are implied in methods based on outflow resistance measurement, and, above all, these methods are hardly suitable for long-time pressure measurement nor for objectivation of dynamic pressure alterations. Described in this paper is the authors' own method, using a wick catheter. It is highly reproducible and is characterised by high measuring dynamics. A comparatively large amount of hardware is required. Evidence is produced to the impact of differentiated conditions of examination on measuring results. Values measured by means of this method in the anterotibial compartment of the lower leg did not normally exceed a pressure level of 2.0 kPa.

摘要

目前尚无普遍接受的测量筋膜间室压力的方法。基于流出阻力测量的方法存在诸多显著缺点,最重要的是,这些方法几乎不适用于长时间压力测量,也不适用于动态压力变化的客观化测量。本文介绍了作者自己使用灯芯导管的方法。该方法具有高度可重复性,并且具有高测量动态性的特点。需要相对大量的硬件设备。文中给出了不同检查条件对测量结果影响的证据。通过该方法在小腿胫前筋膜间室测得的值通常不会超过2.0千帕的压力水平。