Munir Selin, Suzuki Leina, Dixon Michael
Medical Affairs Department, Medacta Australia, Lane Cove, New South Wales, Australia.
Orthopaedic Department, The Sutherland Hospital, Caringbah, New South Wales, Australia.
Arthroplast Today. 2023 Jul 22;22:101157. doi: 10.1016/j.artd.2023.101157. eCollection 2023 Aug.
Collared femoral stems have been considered to reduce the risk of early subsidence over collarless stems. However, with advances in material technology, new surface treatments have been introduced into cementless stem design to enhance primary fixation and long-term stability. This study aims to analyze the early migration behaviors of a proximally coated collarless femoral stem and cementless acetabular component and compare the outcomes with commercially available cementless stems and acetabular cups.
A total of 24 patients (25 hips) undergoing total hip arthroplasty were recruited and followed up for 2 years. All patients received a Masterloc femoral stem (Medacta International SA, Castel San Pietro, Switzerland) and an Mpact acetabular component (Medacta International SA, Castel San Pietro, Switzerland) with tantalum beads embedded during the operation. Radiographs for radiostereometric analysis were taken immediately postsurgery, 6-months, 1-year, and 2-years postoperatively.
The median condition number for this study was 59. The median stem subsidence was -0.08 mm (-2.47 to 0.40) at 2 years. The median cup subsidence was -0.03 mm (-0.38 to 0.57) at 2 years. The migration of the Masterloc stem was less than that of other cementless collarless, as well as collared stems, as reported in literature.
This study has demonstrated the high stability and fixation provided with the use of a collarless cementless stem. The subsidence seen in both this cementless femoral stem and acetabular cup at 2 years postoperative was below the range reported in literature for cementless collarless and collared stems.
带领股骨柄被认为可降低早期下沉风险,优于无领股骨柄。然而,随着材料技术的进步,新型表面处理已被引入非骨水泥型股骨柄设计中,以增强初始固定和长期稳定性。本研究旨在分析近端涂层无领股骨柄和非骨水泥髋臼部件的早期移位行为,并将结果与市售非骨水泥股骨柄和髋臼杯进行比较。
共纳入24例(25髋)接受全髋关节置换术的患者,并随访2年。所有患者均接受了Masterloc股骨柄(瑞士卡斯特尔圣彼得罗的Medacta International SA公司)和Mpact髋臼部件(瑞士卡斯特尔圣彼得罗的Medacta International SA公司),术中植入钽珠。术后立即、术后6个月、1年和2年拍摄用于放射立体测量分析的X线片。
本研究的中位数条件数为59。2年时,股骨柄下沉中位数为-0.08mm(-2.47至0.40)。2年时,髋臼杯下沉中位数为-0.03mm(-0.38至0.57)。如文献报道,Masterloc股骨柄的移位小于其他非骨水泥无领股骨柄以及带领股骨柄。
本研究证明了使用无领非骨水泥股骨柄具有较高的稳定性和固定性。该非骨水泥股骨柄和髋臼杯在术后2年的下沉低于文献报道的非骨水泥无领和带领股骨柄的范围。