Green K, Harman D, Cheeks L
Curr Eye Res. 1986 Aug;5(8):621-3. doi: 10.3109/02713688609015127.
The relationship between the pressure applied to the enucleated human eye using the Honan Intraocular Pressure Reducer (HIPR) and the peak intraocular pressure as a function of initial intraocular pressure has been examined. The peak intraocular pressure is linearly related to the applied HIPR pressure whether the latter is 30, 50, or 75 mm Hg. The slopes relating peak intraocular pressure to initial intraocular pressure at different HIPR settings are parallel. Use of the HIPR at settings greater than 30 mm Hg and an initial IOP of greater than 30 mm Hg could compromise ocular vascular perfusion.
使用河南眼压降低器(HIPR)对摘除的人眼施加的压力与作为初始眼压函数的峰值眼压之间的关系已得到研究。无论施加的HIPR压力是30、50还是75毫米汞柱,峰值眼压都与施加的HIPR压力呈线性相关。在不同的HIPR设置下,将峰值眼压与初始眼压相关联的斜率是平行的。在大于30毫米汞柱的设置下使用HIPR以及初始眼压大于30毫米汞柱可能会损害眼部血管灌注。