Abstract

AIM

During percutaneous coronary intervention (PCI), complicated lesions in the target coronary artery often hinder device delivery. Fluid lubricants have commonly been used to reduce friction between adjacent solid materials in manufacturing, thus achieving smoother action. This experimental study examined whether a contrast medium could function as a fluid lubricant during PCI.

METHODS AND RESULTS

We used two different coronary artery lesion models with distinct complexities made from silicon. Each model was fit into the PCI-simulation system. This laboratory equipment consisted of ordinary PCI instruments and an aorta model from the Valsalva sinus to the descending aorta. A Wolverine™ cutting balloon catheter was advanced through each lesion model via a guide catheter set into the system. The maximum force required to push the catheter through the lesion models was measured while the vessel system was filled with either normal saline or contrast medium. The maximum force required was significantly lower with the contrast medium (1.38 ± 0.21 N in the normal-saline condition vs. 0.92 ± 0.05 N in the contrast-medium condition in the lesion model A, p < 0.001; 1.30 ± 0.07 N in the normal-saline condition vs. 1.14 ± 0.04 N in the contrast-medium condition in the lesion model B, p < 0.001).

CONCLUSIONS

The contrast medium for vessel system filling reduced the force required to push the devices through the lesion models. This contrast medium represents a potential candidate for a liquid lubricant to facilitate device delivery for complicated coronary lesions.

摘要

目的

在经皮冠状动脉介入治疗(PCI)过程中,目标冠状动脉的复杂病变常常会阻碍器械输送。在制造业中,液体润滑剂通常用于减少相邻固体材料之间的摩擦,从而实现更顺畅的操作。本实验研究旨在探讨在PCI过程中造影剂是否可作为液体润滑剂发挥作用。

方法与结果

我们使用了两种由硅制成的具有不同复杂程度的冠状动脉病变模型。每个模型都安装到PCI模拟系统中。该实验室设备由普通的PCI器械和一个从瓦尔萨尔瓦窦到降主动脉的主动脉模型组成。通过系统中设置的引导导管,将Wolverine™切割球囊导管推进穿过每个病变模型。在血管系统充满生理盐水或造影剂的情况下,测量将导管推过病变模型所需的最大力。使用造影剂时所需的最大力显著更低(病变模型A中,生理盐水条件下为1.38±0.21 N,造影剂条件下为0.92±0.05 N,p<0.001;病变模型B中,生理盐水条件下为1.30±0.07 N,造影剂条件下为1.14±0.04 N,p<0.001)。

结论

用于填充血管系统的造影剂降低了将器械推过病变模型所需的力。这种造影剂代表了一种潜在的液体润滑剂候选物,可促进复杂冠状动脉病变的器械输送。