From the Department of Biomedical Engineering, Carnegie Mellon University, Pittsburgh, Pennsylvania.
Department of Surgical Intensive Care Medicine, Nippon Medical School, Tokyo, Japan.
ASAIO J. 2024 Feb 1;70(2):e23-e26. doi: 10.1097/MAT.0000000000002030. Epub 2023 Aug 14.
A new, lightweight (2.3 kg), ambulatory pulmonary assist system (PAS) underwent preliminary evaluation in ambulatory sheep. The PAS was purposefully designed for long-term extracorporeal respiratory support for chronic lung disease and utilizes a novel, small (0.9 m 2 surface area) gas exchanger, the pulmonary assist device, with a modified Heart Assist 5 pump fitting in a small wearable pack. Prototype PAS were attached to two sheep in venovenous configuration for 7 and 14 days, evaluating ability to remain thrombus free; maintain gas exchange and blood flow resistance; avoid biocompatibility-related complications while allowing safe ambulation. The PAS achieved 1.56 L/min of flow at 10.8 kRPM with a 24 Fr cannula in sheep one and 2.0 L/min at 10.5 kRPM with a 28 Fr cannula in sheep 2 without significant change. Both sheep walked freely, demonstrating the first application of truly ambulatory ECMO in sheep. While in vitro testing evaluated PAS oxygen transfer rates of 104.6 ml/min at 2 L/min blood flow, oxygen transfer rates averaged 60.6 ml/min and 70.6 ml/min in studies 1 and 2, due to average hemoglobin concentrations lower than humans (8.9 and 10.5 g/dl, respectively). The presented cases support uncomplicated ambulation using the PAS.
一种新型轻便(2.3 公斤)、可移动的肺辅助系统(PAS)在可移动绵羊中进行了初步评估。该 PAS 是专门为慢性肺部疾病的长期体外呼吸支持而设计的,它采用了一种新型的小型(0.9 平方米表面积)气体交换器,即肺辅助装置,以及一个经过改良的 HeartAssist 5 泵,可安装在一个小型可穿戴包中。原型 PAS 在羊中采用静脉-静脉配置连接了两只羊 7 天和 14 天,评估其保持无血栓的能力;维持气体交换和血流阻力;避免与生物相容性相关的并发症,同时允许安全的移动。在羊 1 中,使用 24Fr 套管 PAS 以 10.8 kRPM 的转速实现 1.56 L/min 的流量,使用 28Fr 套管实现 2.0 L/min 的流量,在羊 2 中以 10.5 kRPM 的转速实现 2.0 L/min 的流量,没有明显变化。两只羊都能自由行走,展示了绵羊中真正可移动 ECMO 的首次应用。尽管体外测试评估了 PAS 在 2 L/min 血流时的氧转移率为 104.6 ml/min,但在研究 1 和 2 中,氧转移率平均为 60.6 ml/min 和 70.6 ml/min,这是由于平均血红蛋白浓度低于人类(分别为 8.9 和 10.5 g/dl)。所呈现的病例支持使用 PAS 进行简单的移动。