Röschmann P, Tischler R
Radiology. 1986 Oct;161(1):251-5. doi: 10.1148/radiology.161.1.3763875.
We describe the design and application of surface coils for magnetic resonance (MR) imaging at high resonance frequencies (85 MHz). Circular, rectangular-frame, and reflector-type surface coils were used in the transmit-and-receive mode. With these coils, the required radio frequency power is reduced by factors of two up to 100 with respect to head and body coils. With the small, circular coils, high-resolution images of a small region of interest can be obtained that are free of foldback and motion artifacts originating outside the field of interest. With the rectangular-frame and reflector coils, large fields of view are also accessible. As examples of applications, single- and multiple-section images of the eye, knee, head and shoulder, and spinal cord are provided.