Suppr超能文献

具有多狭缝扫描X射线束的数字放射成像系统:初步报告。

Digital radiographic imaging system with multiple-slit scanning x-ray beam: preliminary report.

作者信息

Doi K, Fujita H, Ohara K, Ono K, Matsui H, Giger M L, Chan H P

出版信息

Radiology. 1986 Nov;161(2):513-8. doi: 10.1148/radiology.161.2.3763924.

Abstract

A digital imaging system that employs a novel scanning x-ray tube, a multiple-slit assembly (MSA), and an image intensifier (II)-TV digital system is described. Use of the MSA can increase x-ray utilization by a factor of 100 over that obtained with single-slit systems. Scatter and veiling glare can be reduced substantially by synthesizing the final image from a number of multiple-slit images, resulting in improvement in contrast sensitivity. An experimental prototype system consisted of a conventional x-ray tube and an II-TV digital system used in conjunction with mechanical scanning of the MSA. Attenuation curves measured with an aluminum stepwedge showed that scatter and veiling glare were eliminated by this approach. Reconstructed images of phantoms were superior to images obtained with conventional wide-beam exposure. Physical parameters related to design of the scanning x-ray tube and a digital system for acquiring images of the chest are discussed.

摘要

描述了一种数字成像系统,该系统采用新型扫描X射线管、多狭缝组件(MSA)和影像增强器(II)-电视数字系统。与单狭缝系统相比,使用MSA可使X射线利用率提高100倍。通过合成多个多狭缝图像的最终图像,可大幅降低散射和蒙片眩光,从而提高对比度灵敏度。一个实验原型系统由一个传统X射线管和一个与MSA机械扫描结合使用的II-电视数字系统组成。用铝阶梯楔形块测量的衰减曲线表明,这种方法消除了散射和蒙片眩光。假体的重建图像优于传统宽束曝光获得的图像。讨论了与扫描X射线管设计及用于获取胸部图像的数字系统相关的物理参数。