PRCIS

Monitoring visual fields (VFs) through virtual reality devices proved to have good inter-test and test-retest reliability, as well as easy usability, when self-administered by individuals with and without VF defects in a remote setting.

PURPOSE

To assess the reliability of remote, self-administered VF monitoring using a virtual reality VF (VRVF) device in individuals without ocular disease and with stable VF defects.

MATERIALS AND METHODS

Individuals without ocular disease and with stable defects were recruited. All participants had a baseline standard automated perimetry (SAP) test. Participants tested remotely on a VRVF device for 4 weeks (examinations V 1 , V 2 , V 3 , and V 4 ), with the last 3 unassisted. The mean sensitivities of VRVF results were compared with each other and to SAP results for reliability.

RESULTS

A total of 42 eyes from 21 participants were tested on the VRVF device. Participants tested consistently although external factors impacted outcomes. VRVF results were in reasonable agreement with the baseline SAP. Examinations performed by the cohort with stable defects evinced better agreement with SAP examinations (V2, P = 0.79; V3, P = 0.39; V4, P = 0.35) than those reported by the cohort without ocular disease (V2, P = 0.02; V3, P = 0.15; V4, P = 0.22), where the null hypothesis is that the instruments agree. Fixation losses were high and variable in VRVF examinations compared with those of SAP, particularly in certain test takers. Participants considered the device comfortable and easy to use.

CONCLUSIONS

Self-administered, remote VF tests on a VRVF device showed satisfactory test-retest reliability, good inter-test agreement with SAP, and acceptability by its users. External factors may impact at-home testing and age and visual impairment may hinder fixation. Future studies to expand the sample size and understand inconsistencies in fixation losses are recommended.

摘要

PRCIS

在远程环境中,使用虚拟现实设备对有和无视野缺陷的个体进行自我测试时,视野(VF)监测的结果具有良好的组内测试和测试-再测试可靠性,以及易用性。

目的

评估在无眼部疾病和稳定视野缺陷的个体中,使用虚拟现实视野(VRVF)设备进行远程自我管理的 VF 监测的可靠性。

材料和方法

招募无眼部疾病且视野缺陷稳定的个体。所有参与者均进行基线标准自动视野计(SAP)检查。参与者在 VRVF 设备上进行远程测试,为期 4 周(检查 V1、V2、V3 和 V4),最后 3 次无人协助。比较 VRVF 结果的平均敏感度,以评估其与 SAP 结果的可靠性。

结果

共有 21 名参与者的 42 只眼在 VRVF 设备上进行了测试。尽管外部因素会影响结果,但参与者的测试结果始终保持一致。VRVF 结果与基线 SAP 结果具有合理的一致性。与无眼部疾病的队列相比,具有稳定缺陷的队列的检查结果与 SAP 检查结果具有更好的一致性(V2,P=0.79;V3,P=0.39;V4,P=0.35)(无效假设是仪器的结果是一致的)。与 SAP 相比,VRVF 检查中的固视丢失率较高且波动较大,尤其是在某些受检者中。参与者认为该设备舒适且易于使用。

结论

使用 VRVF 设备进行自我管理的远程 VF 测试显示出令人满意的测试-再测试可靠性,与 SAP 测试具有良好的组间一致性,并且得到了用户的认可。外部因素可能会影响家庭测试,年龄和视力障碍可能会影响固视。建议进行未来的研究,以扩大样本量并了解固视损失的不一致性。