The virtual scale endoscope (VSE) is a newly introduced endoscope that helps endoscopists in measuring colorectal polyp size (CPS) during colonoscopy by displaying a virtual scale. This study aimed to determine the usefulness of the VSE for CPS measurement and the educational benefit of using VSE images to improve CPS estimation accuracy.
This study included 42 colorectal polyps in 26 patients treated at Hiroshima University Hospital. In study 1, CPS measured using a VSE before endoscopic mucosal resection was compared with CPS measured on resected specimens, and the agreement between the two measurement methods was evaluated via Bland-Altman analysis. In study 2, 14 endoscopists (5 beginners, 5 intermediates, and 4 experts) took a pre-test to determine the size of 42 polyps. After the pre-test, a lecture on CPS measurement using VSE images was given. One month later, the endoscopists took a post-test to compare CPS accuracy before and after the lecture.
In study 1, Bland-Altman analysis revealed no fixed or proportional errors. The mean bias ±95% limits of agreement (±1.96 standard deviations) of the measurement error was -0.05 ± 0.21 mm, indicating that the agreement between two measurement methods was sufficient. In study 2, the accuracy of CPS measurement was significantly higher among beginners (59.5% vs. 26.7%, p < 0.01) and intermediates (65.2% vs. 44.3%, p < 0.05) in the post-test than in the pre-test.
The VSE accurately measures CPS before resection, and its images are useful teaching tools for beginner and intermediate endoscopists.
虚拟内镜(VSE)是一种新引入的内镜,它通过显示虚拟标尺帮助内镜医生在结肠镜检查时测量结直肠息肉大小(CPS)。本研究旨在确定 VSE 用于 CPS 测量的有用性,以及使用 VSE 图像提高 CPS 估计准确性的教育益处。
本研究纳入了在广岛大学医院治疗的 26 名患者的 42 个结直肠息肉。在研究 1 中,比较了 VSE 测量的内镜黏膜切除前 CPS 与切除标本测量的 CPS,并通过 Bland-Altman 分析评估两种测量方法之间的一致性。在研究 2 中,14 名内镜医生(5 名初学者、5 名中级医生和 4 名专家)进行了预测试以确定 42 个息肉的大小。预测试后,进行了 VSE 图像 CPS 测量的讲座。一个月后,内镜医生进行了后测试,以比较讲座前后的 CPS 准确性。
在研究 1 中,Bland-Altman 分析未显示出固定或比例误差。测量误差的均值 ±95%一致性界限(±1.96 标准差)为-0.05 ± 0.21mm,表明两种测量方法之间的一致性足够。在研究 2 中,初学者(59.5%比 26.7%,p < 0.01)和中级医生(65.2%比 44.3%,p < 0.05)在后测试中的 CPS 测量准确性明显高于前测试。
VSE 可以准确测量切除前的 CPS,其图像是初学者和中级内镜医生的有用教学工具。