In this work, we developed a new copper(II)-selective potentiometric sensor and investigated its surface with scanning electron microscopy (SEM). Besides the surface images of the sensors conditioned in copper(II) solutions, energy-dispersive X-ray (EDX) and mapping studies were carried out. According to the results obtained, it was determined that copper(II) ions adhered to the porous areas on the sensor surface, and that Cu ions showed a wide distribution on the sensor surface in mapping studies. The new sensor had a Nernstian response of 29.3 ± 0.5 mV/decade in the concentration range of 1.0 × 10-1.0 × 10 M and a low detection limit of 8.56 × 10 M. The proposed sensor had fast response time (< 10 s), wide pH working range (5.0-10.0), good repeatability and stability. Finally, the sensor performed the determination of copper(II) ions in various water samples with very high recoveries (96.0-102.0%).
在本工作中,我们开发了一种新型的铜(II)选择性电位传感器,并用扫描电子显微镜(SEM)对其表面进行了研究。除了在铜(II)溶液中调节后的传感器表面图像外,还进行了能量色散X射线(EDX)和映射研究。根据所得结果,确定铜(II)离子附着在传感器表面的多孔区域,并且在映射研究中铜离子在传感器表面呈现出广泛分布。新型传感器在1.0×10⁻¹.0×10⁻⁶M的浓度范围内具有29.3±0.5mV/十倍的能斯特响应,检测限低至8.56×10⁻⁷M。所提出的传感器具有快速响应时间(<10秒)、较宽的pH工作范围(5.0 - 10.0)、良好的重复性和稳定性。最后,该传感器对各种水样中的铜(II)离子进行测定,回收率非常高(96.0 - 102.0%)。