J Clin Monit Comput. 2023 Dec;37(6):1489-1495. doi: 10.1007/s10877-023-01076-4. Epub 2023 Oct 13.
SentriO Oxy™ is a newly available, Food and Drug Administration-approved oxygenation mask system that provides high oxygenation, even on low-flow (5-10 L/min) oxygen. This study aimed to accurately measure the intratracheal fraction of inspired oxygen (FiO) using SentriO Oxy™ masks under relatively low oxygen flow rates. A manikin-ventilator-test lung simulation system was used. We measured FiO at the level of the carina, 5 minutes after applying 45 different respiratory parameter combinations using SentriO Oxy™ masks. Tidal volume (TV) was set to 300, 500, and 700 mL; respiratory rate (RR) was set to 8, 12, 16, 20, and 24 breaths per minute; and oxygen flow rate was set to 6, 8, and 10 L/min. Our hypothesis was that FiO would be proportional to the difference between oxygen flow rate and minute ventilation. FiO measured by smaller TV, lower RR, or higher oxygen flows revealed a significantly higher value, confirming our hypothesis. In addition, using linear regression analysis, we found that TV, RR, and oxygen flow were all significant factors influencing the measured FiO. Our experiment proposed two prediction equations considering the oxygen flow rate, TV, and RR. The results of our study may provide information and prediction of FiO for clinicians to use SentriO Oxy™ masks during sedative anesthetic procedures under low oxygen flow rates.
SentriO Oxy™ 是一种新的、获得美国食品和药物管理局批准的氧合面罩系统,即使在低流量(5-10 L/min)氧气下也能提供高氧合。本研究旨在使用 SentriO Oxy™ 面罩在相对较低的氧气流速下准确测量气管内吸入氧分数(FiO)。使用模拟系统模拟人-呼吸机-肺。我们在使用 SentriO Oxy™ 面罩施加 45 种不同呼吸参数组合 5 分钟后,在隆突水平测量 FiO。潮气量(TV)设置为 300、500 和 700 mL;呼吸频率(RR)设置为 8、12、16、20 和 24 次/分钟;氧气流量设置为 6、8 和 10 L/min。我们的假设是 FiO 与氧气流量和分钟通气量之间的差异成正比。使用较小的 TV、较低的 RR 或更高的氧气流量测量的 FiO 值显示出显著更高的值,证实了我们的假设。此外,通过线性回归分析,我们发现 TV、RR 和氧气流量都是影响测量 FiO 的重要因素。我们的实验考虑了氧气流量、TV 和 RR,提出了两个预测方程。我们的研究结果可能为临床医生在低氧气流量下进行镇静麻醉程序时使用 SentriO Oxy™ 面罩提供 FiO 的信息和预测。