乳腺摄影系统的剂量比较。

Dose comparisons for mammographic systems.

作者信息

Speiser R C, Zanrosso E M, Jeromin L S, Carlson R A

出版信息

Med Phys. 1986 Sep-Oct;13(5):667-73. doi: 10.1118/1.595871.

Abstract

Dose estimates were made for Kodak Min-R screens combined with Kodak Min-R film and Kodak Ortho M film, both with and without a 5:1 Bucky grid; for standard xeroradiographic techniques in negative development mode; and for the new, higher sensitivity xeroradiographic process of the Xerox 175 System. The estimates were derived from exposure versus depth measurements in phantoms made of BR12 breast simulation material using thermoluminescent detectors. A molybdenum target source with molybdenum filtration, at a half-value layer of 0.37-mm Al, was used for the screen-film measurements. All xeroradiographic measurements were made with a tungsten target source with aluminum filtration at half-value layers of 1.5 to 1.56 mm Al. Mean glandular dose estimates for the Min-R screen/Ortho M film combination with Bucky grid and for the new xeroradiographic process were found to be similar. Dose reduction with the new xeroradiographic system was achieved through a more sensitive photoreceptor and more sensitive development, which also improved the unique imaging characteristics of xeroradiography.

摘要

针对柯达Min - R增感屏与柯达Min - R胶片以及柯达正射M胶片组合进行了剂量估算,有5:1滤线栅和无滤线栅两种情况均做了估算;还对负性显影模式下的标准干板X线摄影技术以及施乐175系统新的高灵敏度干板X线摄影工艺进行了剂量估算。这些估算值来自于使用热释光探测器对由BR12乳腺模拟材料制成的模体进行的曝光量与深度测量。在屏 - 片测量中使用了钼靶源并进行钼滤过,半值层为0.37毫米铝。所有干板X线摄影测量均使用钨靶源并进行铝滤过,半值层为1.5至1.56毫米铝。发现带有滤线栅的Min - R增感屏/正射M胶片组合以及新的干板X线摄影工艺的平均腺体剂量估算值相似。新的干板X线摄影系统通过更灵敏的感光体和更灵敏的显影实现了剂量降低,这也改善了干板X线摄影独特的成像特性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验