验证在胫骨组件 phantom 研究中,评估放射立体测量植入物迁移时,暴露内患者运动的影响。

Validation for the effect of intra-exposure patient motion on the assessment of radiostereometric implant migration in a tibial component phantom study.

机构信息

Department of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg, 91054 Erlangen, Germany; Department of Orthopaedic Surgery, Taizhou Second People's Hospital, 225599 Taizhou, PR China.

Department of Orthopaedic Surgery, The First Affiliated Hospital of Gannan Medical University, 341000 Ganzhou, PR China.

出版信息

Orthop Traumatol Surg Res. 2024 Apr;110(2):103776. doi: 10.1016/j.otsr.2023.103776. Epub 2023 Nov 26.

Abstract

BACKGROUND

An increasing number of radiostereometry (RSA) research studies have long-term follow-up implant migration outcomes, which show ascending curves of implant migration with occasionally decreasing migration. After scrutinizing images and RSA scenes related to the alternating curves, we suppose that intra-exposure patient motion may contribute to that. The main purposes of this in vitro study were 1) to identify whether the patient motion in different directions could result in the inaccurate assessment of implant migration, and 2) to figure out which direction(s) accounted for the alternating curves.

HYPOTHESIS

It was hypothesized that the assessments of implant migration would be less precise and accurate than they could be when patient motion occurred, and such motion would contribute to the alternating curves of radiostereometric implant migration.

MATERIALS AND METHODS

A customized phantom, assembled with a tibial component, was designed for simulating intra-exposure patient motion during follow-up RSA examinations. Two different Roentgen tubes were used as the current standard of radiology departments. Radiographs were acquired in a uniplanar technical arrangement. Two defined protocols were conducted: one is to simulate implant migration outcomes at post-op, the early stage (6months), and the later stage (2 to 10years) ; during the later stage, the other is to mimic patient motion by phantom motion in the medial-lateral (x), distal-proximal (y), and anterior-posterior (z) axes.

RESULTS

Phantom motion could result in the inaccurate assessment of implant migration, and translations along the medial-lateral (x) axis were the most influenced by patient motion. Motion along the medial-lateral (x) axis could account for the curves with decreasing migration.

DISCUSSION

Our assessments of implant migration may be less precise and accurate than they could be when intra-exposure patient motion occurs. We probably neglect the importance of 100% simultaneous exposures, and the influence of patient motion on RSA accuracy and data reliability, due to the difficulty in detecting patient (micro)motion. Electronically synchronized exposures of two paired Roentgen tubes are 100% simultaneous for image acquisition, and they are thus highly recommended for the assessment of implant migration in RSA.

TYPE OF STUDY AND LEVEL OF PROOF

not applicable.

摘要

背景

越来越多的放射立体测量术(RSA)研究具有长期随访的植入物迁移结果,这些结果显示植入物迁移呈上升曲线,偶尔会出现迁移减少的情况。在仔细检查与交替曲线相关的图像和 RSA 场景后,我们假设术中患者运动可能对此有影响。本体外研究的主要目的是:1)确定不同方向的患者运动是否会导致植入物迁移评估不准确,以及 2)确定哪个方向导致了交替曲线。

假设

如果发生患者运动,植入物迁移的评估将不如没有运动时准确和精确,并且这种运动会导致放射立体测量植入物迁移的交替曲线。

材料和方法

设计了一个定制的体模,由胫骨部件组装而成,用于模拟随访 RSA 检查期间的术中患者运动。使用两个不同的伦琴管作为放射科的当前标准。使用单平面技术安排获取射线照片。进行了两种定义的方案:一种是模拟术后、早期(6 个月)和晚期(2 至 10 年)的植入物迁移结果;在晚期,另一种是通过体模在内外(x)、远侧近侧(y)和前后(z)轴上的运动来模拟患者运动。

结果

体模运动可能导致植入物迁移评估不准确,并且沿内外(x)轴的平移受患者运动影响最大。沿内外(x)轴的运动可能导致迁移减少的曲线。

讨论

当术中患者运动发生时,我们对植入物迁移的评估可能不如没有运动时准确和精确。由于难以检测患者(微观)运动,我们可能忽略了 100%同步曝光的重要性,以及患者运动对 RSA 准确性和数据可靠性的影响。两个配对伦琴管的电子同步曝光是 100%同步进行图像采集的,因此强烈建议用于 RSA 中植入物迁移的评估。

研究类型和证据水平

不适用。