uMI550数字PET-CT在常规质量控制程序中的操作与性能经验

Operational and Performance Experience with uMI550 Digital PET-CT during Routine Quality Control Procedures.

作者信息

Singh Manoj Kumar, Mohan V Sai Krishna, Kaushik Chanchal

机构信息

Medikabazaar, Technopolis Knowledge Park, Mumbai, Maharashtra, India.

Medicover Cancer Institute, Nellore, Andhra Pradesh, India.

出版信息

World J Nucl Med. 2023 Dec 26;22(4):267-275. doi: 10.1055/s-0043-1777696. eCollection 2023 Dec.

Abstract

The quality control (QC) procedures for positron emission tomography (PET) scanners are covered by National Electrical Manufacturers Association and International Electrotechnical Commission. QC must be carried out at regular intervals according to the specifications of the scanner manufacturer. Daily and weekly QC plays a valuable role in monitoring positron emission tomography (PET) scanner performance changes. This study shares operational and performance experience of QC procedures that do not require a radioactive Ge-68 source to perform daily QC and experience with fluorodeoxyglucose F18 ( F-FDG) as a substitute for germanium-68/sodium-22 (Ge-68/Na-22) source for weekly QC.  This study was performed on an uMI550 digital positron emission tomography-computed tomography (PET-CT) scanner. In this scanner daily QC checks system temperature and humidity, system count rate, data link status, and voltage. QC was performed at the console control, the position of the scanner table was in the home position pulled out from the gantry, and the room was closed during the quick QC. Weekly full QC check items include look-up table drift, energy drift, time-of-flight status, C-map status, temperature and humidity, and voltage. Weekly full QC was performed with a F-FDG source in a rod phantom source.  Over 200 daily QC tests without a radioactive source Ge-68 phantom and 50 full weekly QC tests using a F-FDG rod phantom were performed with this scanner according to the manufacturer's instructions and a test report was generated. No daily QC errors or warnings were observed during this period.  The new approach for the daily PET QC does not expose operators to radiation. This translates into commercial and operational merits with consistent performance and results.  Reduction in radiation exposure to operating staff during QC procedure in PET-CT scanner.

摘要

正电子发射断层扫描(PET)扫描仪的质量控制(QC)程序由美国国家电气制造商协会和国际电工委员会负责。必须根据扫描仪制造商的规格定期进行质量控制。每日和每周的质量控制在监测正电子发射断层扫描(PET)扫描仪性能变化方面发挥着重要作用。本研究分享了无需放射性锗 - 68源进行每日质量控制的质量控制程序的操作和性能经验,以及使用氟脱氧葡萄糖F18(F-FDG)替代锗 - 68/钠 - 22(Ge-68/Na-22)源进行每周质量控制的经验。

本研究是在一台uMI550数字正电子发射断层扫描 - 计算机断层扫描(PET-CT)扫描仪上进行的。在该扫描仪中,每日质量控制检查系统温度和湿度、系统计数率、数据链路状态和电压。质量控制在控制台进行,扫描仪检查床处于从机架拉出的初始位置,快速质量控制期间房间关闭。每周的全面质量控制检查项目包括查找表漂移、能量漂移、飞行时间状态、C图状态、温度和湿度以及电压。每周的全面质量控制使用F-FDG源在棒状体模中进行。

按照制造商的说明,使用该扫描仪进行了200多次无放射性源锗 - 68体模的每日质量控制测试以及50次使用F-FDG棒状体模的每周全面质量控制测试,并生成了测试报告。在此期间未观察到每日质量控制错误或警告。

每日PET质量控制的新方法不会使操作人员受到辐射。这转化为具有一致性能和结果的商业和运营优势。

降低PET-CT扫描仪质量控制程序期间操作人员所受的辐射剂量。