A Positron Emission Tomography/Computed Tomography quality assurance program is necessary to ensure that patients receive optimal imaging and care. We summarize the AAPM Task Group (TG) 126 report on acceptance and quality assurance (QA) testing of PET/CT systems.
TG 126 was charged with developing PET/CT acceptance testing and QA procedures. The TG aimed to develop procedures that would allow for standardized evaluation of existing short-axis cylindrical-bore PET/CT systems in the spirit of NEMA NU 2 standards without requiring specialized phantoms or proprietary software tools.
We outline eight performance evaluations using common phantoms and freely available software whereby the clinical physicist monitors each PET/CT system by comparing periodic Follow-Up Measurements to Baseline Measurements acquired during acceptance testing. For each of the eight evaluations, we also summarize the expected testing time and materials necessary and the recommended pass/fail criteria.
Our report provides a guideline for periodic evaluations of most clinical PET/CT systems that simplifies procedures and requirements outlined by other agencies and will facilitate performance comparisons across vendors, models, and institutions.
正电子发射断层扫描/计算机断层扫描质量保证程序对于确保患者获得最佳成像和护理至关重要。我们总结了美国医学物理学会任务组(TG)126关于PET/CT系统验收和质量保证(QA)测试的报告。
TG 126负责制定PET/CT验收测试和QA程序。该任务组旨在制定程序,以便能够按照NEMA NU 2标准的精神,对现有的短轴圆柱形孔PET/CT系统进行标准化评估,而无需专门的体模或专有软件工具。
我们概述了八项使用通用体模和免费软件的性能评估,临床物理学家通过将定期的随访测量结果与验收测试期间获取的基线测量结果进行比较,来监测每个PET/CT系统。对于这八项评估中的每一项,我们还总结了预期的测试时间和所需材料以及推荐的通过/失败标准。
我们的报告为大多数临床PET/CT系统的定期评估提供了指导方针,简化了其他机构概述的程序和要求,并将促进不同供应商、型号和机构之间的性能比较。