Suppr超能文献

在小儿心脏移植患者家中使用体积吸收微采样和平行反应监测质谱法测定他克莫司血药谷浓度。

Use of volumetric absorptive microsampling and parallel reaction monitoring mass spectrometry for tacrolimus blood trough measurements at home in pediatric heart transplant patients.

作者信息

Zhao Junfang, Setchell Kenneth D R, Zhao Xueheng, Galandi Stephanie, Garr BreAnn N, Gao Zhiqian, Chin Clifford, Stark Shelly, Steele Paul E, Ryan Thomas D

机构信息

Division of Pathology & Laboratory Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.

Department of Pediatrics, University of Cincinnati College of Medicine, Cincinnati, OH, USA.

出版信息

J Mass Spectrom Adv Clin Lab. 2023 Dec 2;31:1-7. doi: 10.1016/j.jmsacl.2023.11.004. eCollection 2024 Jan.

Abstract

BACKGROUND

Measurement of trough levels for calcineurin inhibitors by venipuncture sampling is a mainstay of patient management in solid organ transplant recipients but challenging in pediatric patients. Volumetric Absorptive Microsampling (VAMS) is a patient-friendly, minimally invasive sampling technique to accurately collect blood. An assay for measurement of tacrolimus in blood using VAMS, coupled with parallel reaction monitoring (PRM) mass spectrometry, was validated in pediatric heart transplant patients.

METHODS

Tacrolimus was measured by a newly developed high-resolution PRM assay and compared with low-resolution tandem mass spectrometry (MRM). Dried blood samples were collected from pediatric heart transplant patients (n = 35) using VAMS devices and a satisfaction survey was completed by patients/guardians. Tacrolimus concentrations were compared across whole liquid blood, dried blood spots, and capillary blood, and shipping stability determined.

RESULTS

The PRM assay was linear over a range 1-50 ng/mL, similar to MRM but had greater specificity due to reduced background noise. No significant differences in tacrolimus concentrations were observed between VAMS and venous blood. Tacrolimus dried on VAM tips was stable for 14 days and concentrations were unaffected by postal shipping. The variability in two simultaneously collected at-home patient samples was minimal - average concentration difference was 0.12 ± 0.94 ng/mL ( = 0.6) between paired samples.

CONCLUSION

A high resolution PRM mass spectrometry assay was developed for home-based dried blood collections for therapeutic monitoring of tacrolimus. The advantage of PRM was enhanced specificity and the VAMS devices provided a simple and convenient approach to blood sampling at home in pediatric heart transplant patients.

摘要

背景

通过静脉穿刺采样来测量钙调神经磷酸酶抑制剂的谷浓度是实体器官移植受者患者管理的主要手段,但在儿科患者中具有挑战性。容积吸收微采样(VAMS)是一种对患者友好的微创采样技术,可准确采集血液。一种使用VAMS采集血液并结合平行反应监测(PRM)质谱法测量他克莫司的检测方法在儿科心脏移植患者中得到了验证。

方法

采用新开发的高分辨率PRM检测法测量他克莫司,并与低分辨率串联质谱法(MRM)进行比较。使用VAMS设备从儿科心脏移植患者(n = 35)中采集干血样,并由患者/监护人完成满意度调查。比较了全血、干血斑和毛细血管血中的他克莫司浓度,并确定了运输稳定性。

结果

PRM检测法在1-50 ng/mL范围内呈线性,与MRM相似,但由于背景噪声降低,具有更高的特异性。VAMS和静脉血之间的他克莫司浓度未观察到显著差异。VAM吸头干燥后的他克莫司在14天内稳定,浓度不受邮政运输的影响。同时采集的两份居家患者样本的变异性最小——配对样本之间的平均浓度差异为0.12±0.94 ng/mL(P = 0.6)。

结论

开发了一种高分辨率PRM质谱检测法,用于居家采集干血样以监测他克莫司的治疗效果。PRM的优势在于特异性增强,VAMS设备为儿科心脏移植患者居家采血提供了一种简单便捷的方法。