Kurki T, Smith N T, Head N, Dec-Silver H, Quinn A
J Clin Monit. 1987 Jan;3(1):6-13. doi: 10.1007/BF00770876.
We recorded finger arterial blood pressure (FINAP) in 50 male patients during various types of surgical operations. Three different types of cuffs were used on four fingers of each patient. Measurements were made by the arterial volume-clamp method of Penaz. The FINAP measurements were compared with pressure data obtained ipsilaterally from a radial artery catheter-transducer system (intraarterial pressure [IAP]) to find optimal recording conditions and to document factors affecting FINAP readings. The thumb, with a specially designed cuff, gave the most accurate results. The mean FINAP - IAP difference for the thumb was -4.8 mm Hg for systolic pressure, 1.49 mm Hg for diastolic pressure, and 0.29 mm Hg for mean pressure. The differences were statistically significant for systolic and diastolic pressure but not for mean pressure. The regression slope for thumb systolic FINAP/IAP was 0.979, that for thumb diastolic FINAP/IAP was 0.963, and that for mean thumb FINAP/IAP was 0.996, whereas the intercepts were 7.499 for systolic pressure, 0.802 for diastolic pressure, and 0.083 for mean pressure. The correlation coefficients were 0.945 (systolic), 0.884 (diastolic), and 0.949 (mean). The correlation coefficients with the other fingers ranged from 0.502 to 0.922 for systolic pressure, 0.757 to 0.932 for diastolic pressure, and 0.767 to 0.892 for mean pressure. The slopes for the various finger-cuff combinations ranged from 0.537 to 0.996, and the intercepts ranged from 0.083 to 32.387 from mean pressure. In 3 patients (6%) the FINAP measurement was not possible because of insufficient peripheral circulation. In 9 other patients (18%) the FINAP measurements were not accurate during some periods of time.(ABSTRACT TRUNCATED AT 250 WORDS)
我们在50名男性患者进行各类外科手术期间记录了手指动脉血压(FINAP)。在每位患者的四根手指上使用了三种不同类型的袖带。测量采用Penaz的动脉容积钳夹法。将FINAP测量值与从桡动脉导管 - 换能器系统同侧获取的压力数据(动脉内压[IAP])进行比较,以找到最佳记录条件并记录影响FINAP读数的因素。使用专门设计的袖带,拇指给出的结果最准确。拇指的平均FINAP - IAP差值,收缩压为 -4.8 mmHg,舒张压为1.49 mmHg,平均压为0.29 mmHg。收缩压和舒张压的差异具有统计学意义,而平均压则无统计学意义。拇指收缩期FINAP/IAP的回归斜率为0.979,拇指舒张期FINAP/IAP的回归斜率为0.963,拇指平均FINAP/IAP的回归斜率为0.996,而截距分别为收缩压7.499、舒张压0.802和平均压0.083。相关系数分别为0.945(收缩期)、0.884(舒张期)和0.949(平均压)。与其他手指的相关系数,收缩压范围为0.502至0.922,舒张压范围为0.757至0.932,平均压范围为0.767至0.892。各种手指袖带组合的斜率范围为0.537至0.996,平均压的截距范围为0.083至32.387。3例患者(6%)因外周循环不足无法进行FINAP测量。另外9例患者(18%)在某些时间段内FINAP测量不准确。(摘要截断于250字)