评估智能药丸(一种无线传感器)作为测量腹腔内压力(IAP)的工具。

Assessment of the Smartpill, a Wireless Sensor, as a Measurement Tool for Intra-Abdominal Pressure (IAP).

机构信息

IFSTTAR, Université Gustave Eiffel, Aix-Marseille Université, Laboratoire de Biomécanique Appliquée, Faculté de Médecine, Campus Nord, Boulevard Pierre Dramard, CEDEX 20, 13916 Marseille, France.

Assistance Publique des Hôpitaux de Marseille (APHM), Aix-Marseille Université, Hôpital Nord, Service de Gastro-Entérologie, Chemin des Bourrely, 13015 Marseille, France.

出版信息

Sensors (Basel). 2023 Dec 21;24(1):54. doi: 10.3390/s24010054.

Abstract

The SmartPill, a multisensor ingestible capsule, is marketed for intestinal motility disorders. It includes a pressure sensor, which could be used to study intra-abdominal pressure (IAP) variations. However, the validation data are lacking for this use. An experimental study was conducted on anesthetized pigs with stepwise variations of IAP (from 0 to 15 mmHg by 3 mmHg steps) generated by laparoscopic insufflation. A SmartPill, inserted by endoscopy, provided intragastric pressure data. These data were compensated to take into account the intrabdominal temperature. They were compared to the pressure recorded by intragastric (IG) and intraperitoneal (IP) wired sensors by statistical Spearman and Bland-Altmann analysis. More than 4500 pressure values for each sensor were generated on two animals. The IG pressure values obtained with the SmartPill were correlated with the IG pressure values obtained with the wired sensor (respectively, Spearman ρ coefficients 0.90 ± 0.08 and 0.72 ± 0.25; bias of -28 ± -0.3 mmHg and -29.2 ± 0.5 mmHg for pigs 1 and 2). The intragastric SmartPill values were also correlated with the IAP measured intra-peritoneally (respectively, Spearman ρ coefficients 0.49 ± 0.18 and 0.57 ± 0.30; bias of -29 ± 1 mmHg and -31 ± 0.7 mmHg for pigs 1 and 2). The SmartPill is a wireless and painless sensor that appears to correctly monitor IAP variations.

摘要

智能胶囊,一种多传感器可摄入胶囊,市场定位于治疗肠道运动障碍。它包含一个压力传感器,可用于研究腹腔内压力(IAP)的变化。然而,这种用途缺乏验证数据。在腹腔镜充气产生的逐步 IAP 变化(从 0 到 15mmHg,每 3mmHg 一步)的麻醉猪上进行了一项实验研究。通过内窥镜插入智能胶囊,提供胃内压力数据。这些数据通过统计 Spearman 和 Bland-Altmann 分析进行补偿,以考虑到腹内温度。将它们与胃内(IG)和腹膜内(IP)有线传感器记录的压力进行比较。两只动物产生了超过 4500 个每个传感器的压力值。智能胶囊获得的 IG 压力值与有线传感器获得的 IG 压力值相关(分别为 Spearman ρ系数 0.90 ± 0.08 和 0.72 ± 0.25;猪 1 和 2 的偏差分别为-28 ± -0.3mmHg 和-29.2 ± 0.5mmHg)。胃内智能胶囊值也与腹膜内测量的 IAP 相关(分别为 Spearman ρ系数 0.49 ± 0.18 和 0.57 ± 0.30;猪 1 和 2 的偏差分别为-29 ± 1mmHg 和-31 ± 0.7mmHg)。智能胶囊是一种无线且无痛的传感器,似乎可以正确监测 IAP 的变化。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4041/10781213/907ebc501989/sensors-24-00054-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验