间歇性导尿管引起的微生物转移:具有可变保护特性的导尿管中微生物转移的体外评估。

Microbial Transfer by Intermittent Catheters: An In Vitro Evaluation of Microbial Transfer in Catheter With Variable Protective Features.

机构信息

Nicole L. Dean, BS, Hollister Inc, Libertyville, Illinois.

James Gras, AS, iFyber LLC, Ithaca, New York.

出版信息

J Wound Ostomy Continence Nurs. 2024;51(1):66-73. doi: 10.1097/WON.0000000000001042.

Abstract

PURPOSE

The purpose of this study was to evaluate the effects of various protective features (eg, catheter cap, introducer tip, and catheter sleeve) of hydrophilic intermittent catheters against contamination with urinary tract infection-associated microorganisms using an in vitro model.

DESIGN

An in vitro study of microbial transfer.

MATERIALS AND METHODS

Gloves were contaminated with uropathogenic microorganisms and used to simulate intermittent catheterization of male anatomical models with and without the protective features present in 5 commercially available hydrophilic catheters. Using this contaminated touch transfer method, both the meatus of the sterile male anatomical models and sterile surgical gloves of an operator were inoculated with a high level of microorganisms (107 and 109 colony-forming units [CFU], respectively). The operator then performed catheterization of the anatomical model. The most relevant segments of the catheter were sampled, and the level of microbial transfer and catheter contamination was quantified. Results from experimental and sample replicates from the 3 microbial species and 5 catheters (sleeved and unsleeved) were analyzed by pair-wise t tests and analysis of variance.

RESULTS

Of the 5 commercially available sleeved intermittent catheters evaluated in this study, use of catheters with multiple protective components (ring cap, introducer tip, and catheter sleeve) resulted in significant improvement in protection against contamination with a 25- to 2500-fold lower level of microbial contamination (C1 segment) across all species as compared to catheters protected with only sleeves or un-sleeved catheters.

CONCLUSIONS

The combination of a ring cap, protective introducer tip, and protective sleeve provides additional protection when compared to sleeve alone from transferring microbial contamination from the meatus to the advancing catheter. Additional research is needed to determine whether these design features result in fewer urinary tract infections among intermittent catheter users.

摘要

目的

本研究旨在通过体外模型评估亲水间歇性导尿管的各种保护特性(如导管帽、导入器尖端和导管套)对尿路感染相关微生物污染的影响。

设计

微生物转移的体外研究。

材料和方法

手套被污染的尿路致病性微生物,并用模拟间歇导尿与和没有目前在 5 种市售亲水导管的保护功能的男性解剖模型。使用这种污染的触摸转移方法,无菌男性解剖模型的尿道口和无菌手术手套的操作者都接种了高水平的微生物(分别为 107 和 109 菌落形成单位 [CFU])。然后操作者对解剖模型进行导尿。对导管的最相关部分进行采样,并量化微生物转移和导管污染的程度。通过配对 t 检验和方差分析对来自 3 种微生物和 5 种导管(带套和不带套)的实验和样本重复的结果进行分析。

结果

在所评估的 5 种市售带套间歇性导尿管中,与仅带套或不带套的导尿管相比,使用具有多个保护部件(环帽、导入器尖端和导管套)的导尿管可显著降低微生物污染程度(C1 段),降低 25-2500 倍。

结论

与仅带套的导尿管相比,环帽、保护导入器尖端和保护套的组合可在将微生物污染从尿道口转移到推进的导管时提供额外的保护。需要进一步研究这些设计特点是否会导致间歇性导尿使用者的尿路感染减少。