Suppr超能文献

二维液相色谱-双内标同位素稀释串联质谱法校正血清肌酐定量中的肌酸-肌酐转换。

Correction of creatine-creatinine conversion during serum creatinine quantification by two-dimensional liquid chromatography and double-spike isotope dilution tandem mass spectrometry.

机构信息

Department of Physical and Analytical Chemistry, University of Oviedo, Avenida Julián Clavería, 8, 33006 Oviedo, Spain.

Department of Physical and Analytical Chemistry, University of Oviedo, Avenida Julián Clavería, 8, 33006 Oviedo, Spain.

出版信息

Clin Chim Acta. 2024 Feb 1;554:117778. doi: 10.1016/j.cca.2024.117778. Epub 2024 Jan 14.

Abstract

BACKGROUND AND AIMS

Development of a candidate reference method based on bidimensional liquid chromatography coupled to ESI-MS/MS and double spike isotope dilution for serum creatinine quantification capable of correcting for creatinine-creatine interconversion during sample pretreatment. Study of the impact of the creatine-creatinine interconversion during the analysis of human serum samples.

MATERIALS AND METHODS

C-creatinine and C-creatine are added to the serum sample. Separation carried out by bidimensional liquid chromatography combining reversed phase and a strong cation exchange chromatography. The heart cut, containing creatine and creatinine, is automatically transferred to the second dimension. Quantification carried out by double spike isotope dilution tandem MS/MS.

RESULTS

Minimization of spectral interferences and ion suppression due to matrix effects while increasing sample throughput compared to the direct coupling of cation exchange chromatography to the ESI source. Trueness of the method studied with the satisfactory analysis of two certified reference materials. Satisfactory intra- and inter-day precisions obtained analysing a serum pool and control sera. Analysis of 93 serum samples revealed negligible interconversions with no correlation with creatine levels.

CONCLUSIONS

The method provides adequate analytical figures of merit for serum creatinine determination according to CSLI guidelines. Negligible creatine-creatinine interconversion is promoted with the applied sample preparation procedure.

摘要

背景和目的

建立一种基于二维液相色谱-电喷雾串联质谱联用和双内标同位素稀释法的候选参考方法,用于血清肌酐的定量分析,该方法能够校正样品预处理过程中肌酐-肌酸的相互转化。研究分析人血清样品过程中肌酐-肌酸相互转化的影响。

材料和方法

在血清样品中加入 C-肌酐和 C-肌酸。通过二维液相色谱法进行分离,结合反相和强阳离子交换色谱。含有肌酸和肌酐的心切产物自动转移到第二维。采用双内标同位素稀释串联质谱法进行定量分析。

结果

与阳离子交换色谱直接与 ESI 源耦合相比,该方法通过基质效应最小化了光谱干扰和离子抑制,同时提高了样品通量。方法的准确性通过对两种有证参考物质的满意分析进行研究。通过分析血清池和对照血清获得了满意的日内和日间精密度。分析 93 份血清样品发现,肌酐-肌酸的相互转化可忽略不计,且与肌酸水平无关。

结论

该方法根据 CSLI 指南提供了足够的血清肌酐测定分析性能。应用的样品制备程序促进了可忽略不计的肌酐-肌酸相互转化。