Howland H C, Sayles N, Cacciotti C, Howland M
Am J Optom Physiol Opt. 1987 Feb;64(2):114-22. doi: 10.1097/00006324-198702000-00007.
A simple pointspread retinoscope and method of retinoscopy are described which use optical principles similar to those of photoretinoscopy or static photographic skiascopy. In this instrument the degree of defocus of an eye is estimated from the disappearance of the bright fundal reflex or "crescent" in the eye of the subject when the distance between the light source and the eye of the observer is increased systematically. The instrument is particularly suitable for large scale screening efforts because it is easy to use and can be constructed from inexpensive materials. The pointspread retinoscope was calibrated against an artificial eye of known defocus. Additionally, results from pointspread retinoscopy were compared against meridional photorefractive data from a sample of infants and young children, and a significant correlation was found to exist between measurements made by the two methods.
本文描述了一种简单的点扩散视网膜检影镜及视网膜检影方法,其运用的光学原理与视网膜电图检影法或静态摄影视网膜镜检法相似。在该仪器中,当系统地增加光源与观察者眼睛之间的距离时,通过观察被检者眼睛中明亮的眼底反射或“新月形”的消失来估计眼睛的散焦程度。该仪器特别适用于大规模筛查工作,因为它易于使用且可以用廉价材料制成。点扩散视网膜检影镜已针对已知散焦的人工眼进行了校准。此外,还将点扩散视网膜检影的结果与一组婴幼儿的子午线屈光数据进行了比较,发现两种方法的测量结果之间存在显著相关性。