Di Staso Federico, Gaudio Paul A
Connecticut Uveitis Foundation, Ocular Immunology, West Hartford, United States.
J Curr Glaucoma Pract. 2023 Oct-Dec;17(4):197-204. doi: 10.5005/jp-journals-10078-1421.
To evaluate the Falck Medical Applanation Tonometer 1 (FMAT-1) as a device for measuring intraocular pressure (IOP) and outflow facility (OF) in patients with uveitis after periocular triamcinolone injection in one eye.
Retrospective observational comparison. IOP and OF measurements were recorded and analyzed, comparing Goldmann tonometry and FMAT-1 readings. Records were reviewed for adult patients with uveitis who had undergone recent periocular triamcinolone injection in one eye.
A total of 19 patients' records were evaluated. In treated eyes, median IOPs (in mm Hg) by Goldmann and FMAT-1 were 21 and 21.0, respectively, and mean IOPs were 22.21 and 21.15, respectively. In untreated eyes, median IOPs by Goldmann and FMAT-1 were 15 and 15.7, respectively, and mean IOPs were 15.52 and 15.31, respectively. Median OF (in µL/mm Hg) in treated and untreated eyes was 0.17 and 0.22, respectively, and mean values were 0.17 and 0.23, respectively. Pearson correlation showed an inverse relationship between IOP and OF in treated eyes, and linear regression analysis showed that IOP strongly predicted outflow in this group.
The FMAT-1 generates IOP results substantially similar to Goldmann and OF readings with greater ease than prior tonography devices.
The mechanism of corticosteroid-induced IOP elevation appears to be a reduction in aqueous OF, and its measurement is tonography. This technique has traditionally been cumbersome. FMAT-1 is a newly launched instrument that measures IOP and OF at the slit lamp simultaneously in a few seconds.
Staso FD, Gaudio PA. Aqueous Outflow Facility after Periocular Triamcinolone Injection: A Preliminary Evaluation of the Falck Medical Applanation Tonometer. J Curr Glaucoma Pract 2023;17(4):197-204.
评估Falck眼压计1型(FMAT - 1)作为一种测量单眼球周注射曲安奈德后葡萄膜炎患者眼压(IOP)和房水流出易度(OF)的设备。
回顾性观察比较。记录并分析眼压和房水流出易度测量值,比较Goldmann眼压测量法和FMAT - 1的读数。对近期单眼接受球周注射曲安奈德的葡萄膜炎成年患者的记录进行回顾。
共评估了19例患者的记录。在治疗眼,Goldmann眼压计和FMAT - 1测量的眼压中位数(单位:mmHg)分别为21和21.0,平均眼压分别为22.21和21.15。在未治疗眼,Goldmann眼压计和FMAT - 1测量的眼压中位数分别为15和15.7,平均眼压分别为15.52和15.31。治疗眼和未治疗眼的房水流出易度中位数(单位:µL/mm Hg)分别为0.17和0.22,平均值分别为0.17和0.23。Pearson相关性分析显示治疗眼中眼压与房水流出易度呈负相关,线性回归分析表明眼压强烈预测该组的房水流出。
FMAT - 1产生的眼压结果与Goldmann眼压计基本相似,且测量房水流出易度的读数与之前的眼压描记设备相比更为简便。
皮质类固醇诱导眼压升高的机制似乎是房水流出易度降低,其测量方法是眼压描记法。传统上这项技术很繁琐。FMAT - 1是一种新推出的仪器,可在裂隙灯处几秒钟内同时测量眼压和房水流出易度。
Staso FD, Gaudio PA. 球周注射曲安奈德后的房水流出易度:Falck眼压计的初步评估。《当代青光眼实践杂志》2023;17(4):197 - 204。