一种新型即时检测试验同时检测抗转谷氨酰胺酶 IgA 和抗脱酰胺麦胶 IgG 抗体的诊断准确性。

Diagnostic accuracy of a novel point-of-care test for simultaneous detection of anti-transglutaminase IgA and anti-deamidated gliadin IgG antibodies.

机构信息

Immunology and Allergy Unit, Santa Maria degli Angeli Hospital, Pordenone, Italy.

PRIMA Lab SA Laboratories, Balerna, Switzerland.

出版信息

J Clin Lab Anal. 2024 Feb;38(3):e25003. doi: 10.1002/jcla.25003. Epub 2024 Jan 30.

Abstract

BACKGROUND

Point-of-care tests (POCTs) may have a role in detecting undiagnosed cases of Celiac disease (CD). We assessed the diagnostic accuracy of a novel POCT, compared with the conventional serological methods, for simultaneous anti-transglutaminase (tTG) IgA and anti-deamidated gliadin (DGP) IgG antibody detection. Furthermore, we evaluated the effect of different biological matrices (whole blood and serum) on test performance.

METHODS

Serum and whole blood from celiac or suspected celiac patients who underwent duodenal biopsy were assayed for the presence of anti-tTG IgA and anti-DGP IgG both with the reference standard methods (Thermo Fisher Scientific, Uppsala, Sweden) and with the POCT (PRIMA Lab SA, Balerna, Switzerland).

RESULTS

266 sera (101 negative and 165 positive) and 60 whole blood samples (34 positive and 26 negative) were included in the study. POCT for anti-DGP IgG showed a sensitivity of 84.3% and a specificity of 90.1%, with positive (PPV) and negative predictive values (NPV) of 91.07% and 82.73%. POCT for anti-tTG IgA showed a sensitivity of 98.31% and a specificity of 98.02%, with a PPV and NPV of 98.31% and 98.02%. Test accuracies were 86.94% and 98.17%, respectively. The agreement of the results between the two different matrices showed a strong correlation rate: 95% for anti-DGP IgG and 100% for anti-tTG IgA.

CONCLUSION

The anti-tTG IgA/anti-DGP IgG-based POCT showed good diagnostic accuracy with comparable sensitivities and specificities to reference standard methods in detecting CD in symptomatic patients and could be considered as a mass screening test before referring to conventional serology.

摘要

背景

即时检验(POCT)可能在检测未确诊的乳糜泻(CD)病例方面发挥作用。我们评估了一种新型 POCT 的诊断准确性,与传统的血清学方法相比,用于同时检测抗转谷氨酰胺酶(tTG)IgA 和抗脱酰胺麦胶蛋白(DGP)IgG 抗体。此外,我们评估了不同生物基质(全血和血清)对测试性能的影响。

方法

对接受十二指肠活检的乳糜泻或疑似乳糜泻患者的血清和全血进行抗 tTG IgA 和抗 DGP IgG 的检测,同时使用参考标准方法(瑞典乌普萨拉的 Thermo Fisher Scientific)和 POCT(瑞士巴尔纳的 PRIMA Lab SA)进行检测。

结果

本研究共纳入 266 份血清(101 份阴性和 165 份阳性)和 60 份全血样本(34 份阳性和 26 份阴性)。POCT 检测抗 DGP IgG 的敏感性为 84.3%,特异性为 90.1%,阳性预测值(PPV)和阴性预测值(NPV)分别为 91.07%和 82.73%。POCT 检测抗 tTG IgA 的敏感性为 98.31%,特异性为 98.02%,PPV 和 NPV 分别为 98.31%和 98.02%。测试准确率分别为 86.94%和 98.17%。两种不同基质之间的结果一致性显示出很强的相关率:抗 DGP IgG 为 95%,抗 tTG IgA 为 100%。

结论

基于抗 tTG IgA/抗 DGP IgG 的 POCT 在检测有症状患者的 CD 方面具有良好的诊断准确性,与参考标准方法相比,其敏感性和特异性相当,可考虑作为常规血清学检查前的大规模筛查试验。