Mascini M, Mazzei F, Moscone D, Calabrese G, Benedetti M M
Clin Chem. 1987 Apr;33(4):591-3.
Glucose, lactate, and pyruvate can be continuously measured in whole blood by using three extracorporeal electrochemical biosensors. We fixed two newly developed sensors, for lactate and pyruvate, in series with an endocrine artificial pancreas ("Betalike") without affecting its performance. These new sensors for use with whole blood should be able to clarify the fate of the intermediary metabolites of glucose in diabetic patients receiving infusions of drugs or insulin.
通过使用三种体外电化学生物传感器,可以连续测量全血中的葡萄糖、乳酸和丙酮酸。我们将两个新开发的用于检测乳酸和丙酮酸的传感器与一个内分泌人工胰腺(“类β细胞”)串联固定,且不影响其性能。这些用于全血检测的新型传感器应能够阐明接受药物或胰岛素输注的糖尿病患者体内葡萄糖中间代谢产物的去向。