Suppr超能文献

两步回溯误差校准消除相干扫描干涉术中的倾斜模糊度。

Two-step retrace error calibration removing tilt ambiguity in coherence scanning interferometry.

作者信息

Huang Lei, Wang Tianyi, Austin Corey, Kim Daewook, Idir Mourad

出版信息

Opt Lett. 2024 Feb 1;49(3):590-593. doi: 10.1364/OL.510943.

Abstract

In interferometry measurement, the retrace error often limits its high-precision metrology applications. Retrace error calibration with tilted flats can give a relation between the retrace error and the introduced tilt angles, but there is still an ambiguity between the introduced tilt angles and the tilt terms in the created retrace error. We propose a novel, to the best of our knowledge, two-step calibration method to resolve this tilt ambiguity. It involves additional measurements of spherical mirror(s) with known curvature(s). The experiment shows that the curvature deviation due to the tilt ambiguity can be significantly reduced after applying the proposed method.

摘要

在干涉测量中,回程误差常常限制其高精度计量应用。用倾斜平面进行回程误差校准可以给出回程误差与引入倾斜角之间的关系,但在引入的倾斜角与所产生的回程误差中的倾斜项之间仍存在模糊性。据我们所知,我们提出了一种新颖的两步校准方法来解决这种倾斜模糊性。它涉及对具有已知曲率的球面镜进行额外测量。实验表明,应用所提出的方法后,由于倾斜模糊性导致的曲率偏差可显著降低。