Suppr超能文献

保留股骨部件的钛套管新型陶瓷头:500 多例翻修全髋关节置换术的结果。

New Ceramic Heads With Titanium Sleeves on Retained Femoral Components: Results of Over 500 Revision Total Hip Arthroplasties.

机构信息

Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota.

出版信息

J Arthroplasty. 2024 Sep;39(9S1):S183-S187. doi: 10.1016/j.arth.2024.01.045. Epub 2024 Feb 7.

Abstract

BACKGROUND

Ceramic heads are frequently combined with titanium sleeves in revision total hip arthroplasties (THAs), ostensibly to protect the ceramic head from existing damage to the retained trunnion. Although widely adopted, data on the performance and safety of this construct are minimal. The purpose of this study was to describe implant survivorships, radiographic results, and clinical outcomes of patients who underwent revision THA with a ceramic head and titanium sleeve on a retained femoral component.

METHODS

We identified 516 revision THAs with femoral component retention (328 acetabular-only revisions and 188 bearing surface exchanges) treated with a new ceramic head and titanium sleeve between 2000 and 2020. Mean age at revision was 64 years, 56% were women, and mean body mass index was 30. The indications for revision THA were adverse local tissue reaction (25%), acetabular loosening (24%), dislocation (17%), infection (5%), and other (29%). Kaplan-Meier survivorships were analyzed, radiographs reviewed, and Harris Hip Scores evaluated. Mean follow-up was 4 years (range, 2 to 10).

RESULTS

There were no reoperations or failures for ceramic head fracture, taper corrosion, or head/sleeve disengagement. The 10-year survivorship free of any re-revision was 85%. Indications for the 57 re-revisions included dislocation (33), infection (13), acetabular component loosening (7), periprosthetic fracture (2), psoas impingement (1), and sciatic nerve irritation (1). The 10-year survivorship free of any reoperation was 82%. There were an additional 14 reoperations. Radiographically, 1.9% had progressive femoral radiolucent lines, and 4.7% had progressive acetabular radiolucent lines. Mean Harris Hip Score was 81 at 2 years.

CONCLUSIONS

New ceramic heads with titanium sleeves in revision THAs with retained femoral components were durable and reliable with no cases of ceramic head fracture or taper complications at mean 4-year follow-up, including those revised for adverse local tissue reaction.

LEVEL OF EVIDENCE

IV.

摘要

背景

陶瓷头经常与钛套筒结合使用于翻修全髋关节置换术(THA)中,表面上是为了保护陶瓷头免受现有固定杯臼损伤。尽管这种组合被广泛采用,但关于这种结构的性能和安全性的数据很少。本研究的目的是描述在保留股骨部件的情况下,使用陶瓷头和钛套筒进行翻修 THA 的患者的植入物存活率、影像学结果和临床结果。

方法

我们在 2000 年至 2020 年间,共发现了 516 例股骨部件保留的翻修 THA(328 例单纯髋臼翻修,188 例关节面置换),使用了新的陶瓷头和钛套筒。翻修时的平均年龄为 64 岁,56%为女性,平均体重指数为 30。翻修 THA 的指征为局部组织不良反应(25%)、髋臼松动(24%)、脱位(17%)、感染(5%)和其他(29%)。分析了 Kaplan-Meier 存活率,对 X 线片进行了复查,并评估了 Harris 髋关节评分。平均随访时间为 4 年(范围 2 至 10 年)。

结果

没有陶瓷头骨折、锥度腐蚀或头/套筒分离的再手术或失败。无任何再翻修的 10 年存活率为 85%。57 例再翻修的指征包括脱位(33 例)、感染(13 例)、髋臼部件松动(7 例)、假体周围骨折(2 例)、腰大肌撞击(1 例)和坐骨神经刺激(1 例)。无任何再手术的 10 年存活率为 82%。还有 14 例再手术。影像学上,1.9%有进行性股骨透亮线,4.7%有进行性髋臼透亮线。2 年时平均 Harris 髋关节评分为 81 分。

结论

在保留股骨部件的翻修 THA 中,新型陶瓷头与钛套筒结合使用具有耐用性和可靠性,在平均 4 年的随访中,没有陶瓷头骨折或锥度并发症的病例,包括因局部组织不良反应而翻修的病例。

证据水平

IV 级。