Magn Reson Imaging. 2024 Jul;110:17-22. doi: 10.1016/j.mri.2024.03.003. Epub 2024 Mar 5.
To compare the image quality of multiplexed sensitivity-encoding diffusion-weighted imaging (MUSE-DWI) and single-shot echo-planar imaging (SS-EPI-DWI) techniques in uterine MRI.
Eighty-eight eligible patients underwent MUSE-DWI and SS-EPI-DWI examinations simultaneously using a 3.0 T MRI system. Two radiologists independently performed quantitative and qualitative analysis of the two groups of images using a double-blind method. The weighted Kappa test was used to evaluate the interobserver agreement. Wilcoxon's rank sum test was used for qualitative parameters, and paired t-test was used for quantitative parameters. Spearman rank correlation analysis was used to obtained correlation between pathological results and mean apparent diffusion coefficient (ADC) value.
The qualitative and quantitative analysis of the images by the two radiologists were in good or excellent agreement, with weighted kappa value ranging from 0.636 to 0.981. The scores of total subjective image quality (15.4 ± 0.99) and signal-to-noise ratio (158.99 ± 60.71) of MUSE-DWI were significantly higher than those of SS-EPI-DWI (12.93 ± 1.62 P < 0.001; 130.23 ± 48.29 P < 0.05). It effectively reduced image distortion and artifact, and had better lesion conspicuity. There was no significant difference in contrast-to-noise ratio score and average ADC values between the two DWI sequences. The average ADC values of the two DWI sequences were highest in the normal uterus group and lowest in the endometrial cancer group, with statistically significant differences among groups (P < 0.01). In addition, the average ADC values of the two DWI sequences were negatively correlated with the type of lesions, decreasing with the malignancy of the lesions (r = -0.805 P < 0.01, r = -0.815 P < 0.01).
Compared to SS-EPI-DWI, MUSE-DWI can significantly reduce distortion, artifacts, and fuzziness in MRI of uterine lesions, which is more conducive to lesion detection.
比较多重敏感编码扩散加权成像(MUSE-DWI)和单次激发回波平面成像(SS-EPI-DWI)技术在子宫 MRI 中的图像质量。
88 例符合条件的患者同时使用 3.0T MRI 系统进行 MUSE-DWI 和 SS-EPI-DWI 检查。两位放射科医生使用双盲法对两组图像进行定量和定性分析。采用加权 Kappa 检验评价观察者间的一致性。对定性参数采用 Wilcoxon 秩和检验,对定量参数采用配对 t 检验。采用 Spearman 秩相关分析获得病理结果与平均表观扩散系数(ADC)值之间的相关性。
两位放射科医生对图像的定性和定量分析一致性较好,加权 Kappa 值范围为 0.636 至 0.981。MUSE-DWI 的总主观图像质量评分(15.4±0.99)和信噪比(158.99±60.71)明显高于 SS-EPI-DWI(12.93±1.62,P<0.001;130.23±48.29,P<0.05)。它有效减少了图像扭曲和伪影,提高了病变的显示能力。两种 DWI 序列的对比噪声比评分和平均 ADC 值无显著差异。两种 DWI 序列的平均 ADC 值在正常子宫组中最高,在子宫内膜癌组中最低,组间差异有统计学意义(P<0.01)。此外,两种 DWI 序列的平均 ADC 值与病变类型呈负相关,随着病变恶性程度的增加而降低(r=-0.805,P<0.01,r=-0.815,P<0.01)。
与 SS-EPI-DWI 相比,MUSE-DWI 可显著减少子宫病变 MRI 中的扭曲、伪影和模糊,更有利于病灶检出。