van Aswegen A, Kleynhans P H, Lötter M G, Minnaar P C, Iturralde M
S Afr Med J. 1979 Jul 14;56(2):50-3.
A non-invasive method for evaluation of regional and global ventricular function by ECG-gated radionuclide blood pool imaging was developed. Cinematic image display is a sensitive method for evaluating wall motion to detect regional ventricular abnormalities and chamber size. A value of r = 0,86 and 0,85 was found for both fixed and variable area radionuclide methods when these ejection fraction values were correlated with cardio-angiographic ejection fraction values in normal subjects and in patients with myocardial infarction. The fixed area method underestimated the ejection fraction in patients with mitra valve stenosis, but this problem can be overcome with the variable area method. The technique developed cannot be used in patients with severe arrhythmias (r = 0,49 and 0,44).
开发了一种通过心电图门控放射性核素血池成像评估局部和整体心室功能的非侵入性方法。电影图像显示是评估室壁运动以检测局部心室异常和腔室大小的敏感方法。当正常受试者和心肌梗死患者的这些射血分数值与心血管造影射血分数值相关时,固定面积和可变面积放射性核素方法的r值分别为0.86和0.85。固定面积法低估了二尖瓣狭窄患者的射血分数,但可变面积法可克服这一问题。所开发的技术不能用于严重心律失常患者(r = 0.49和0.44)。