一名广泛ST段抬高型心肌梗死患者接受了紧急Impella CP®辅助非体外循环冠状动脉搭桥术,未出现机械并发症。

A patient with extensive ST elevation myocardial infarction who underwent emergent Impella CP® assisted off-pump coronary artery bypass without mechanical complication.

作者信息

Machii Yojiro, Sezai Akira, Hayashi Yuki, Harada Atsushi, Kamata Keita, Eguchi Naoki, Tanaka Masashi

机构信息

Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan.

出版信息

J Cardiol Cases. 2023 Dec 7;29(3):104-107. doi: 10.1016/j.jccase.2023.11.008. eCollection 2024 Mar.

Abstract

UNLABELLED

Impella (Abiomed Inc., Danvers, MA, USA) assisted off-pump coronary artery bypass has been increasingly reported in recent years. However, there have been no reports of the procedures performed for acute myocardial infarction in which the patient is hemodynamically unstable. We report a case of a 73-year-old man with cardiogenic shock due to extensive ST elevation acute myocardial infarction that worsened despite Impella CP® support. Because of the fragile myocardium in the acute phase of myocardial infarction, Impella assisted off-pump coronary artery bypass graft causes a high risk of myocardial injury, but we were able to safely perform the procedure by ingenious techniques.

LEARNING OBJECTIVE

When performing Impella-assisted off-pump coronary artery bypass for extensive acute myocardial infarction patients, the high risk of mechanical complications due to myocardial fragility must be considered. The position of Impella should be carefully monitored intraoperatively, and elevation of cardiac apex should be kept to a minimum to prevent myocardial damage caused by Impella.

摘要

未标注

近年来,越来越多的报道称应用Impella(美国马萨诸塞州丹弗斯市的Abiomed公司生产)辅助非体外循环冠状动脉搭桥术。然而,尚无关于对血流动力学不稳定的急性心肌梗死患者实施该手术的报道。我们报告一例73岁男性患者,因广泛ST段抬高型急性心肌梗死导致心源性休克,尽管使用了Impella CP®支持,病情仍恶化。由于心肌梗死急性期心肌脆弱,Impella辅助非体外循环冠状动脉搭桥术会导致心肌损伤的高风险,但我们通过巧妙的技术成功安全地实施了该手术。

学习目标

在对广泛急性心肌梗死患者实施Impella辅助非体外循环冠状动脉搭桥术时,必须考虑到心肌脆弱导致机械并发症的高风险。术中应仔细监测Impella的位置,并将心尖抬高保持在最低限度,以防止Impella造成心肌损伤。

相似文献

2
Off-pump myocardial revascularization with Impella 5.5-assisted for cardiogenic shock.
J Card Surg. 2021 Oct;36(10):3898-3900. doi: 10.1111/jocs.15778. Epub 2021 Jun 30.
3
Percutaneous Transvalvular Microaxial Flow Pump Support in Cardiology.
Circulation. 2022 Apr 19;145(16):1254-1284. doi: 10.1161/CIRCULATIONAHA.121.058229. Epub 2022 Apr 18.
4
Thirty-five day Impella 5.0 support via right axillary side graft cannulation for acute cardiogenic shock.
Innovations (Phila). 2013 Jul-Aug;8(4):307-9. doi: 10.1097/IMI.0000000000000009.
5
One-week Impella CP support for papillary muscle rupture as a bridge to surgery: a case report.
Eur Heart J Case Rep. 2023 Jul 3;7(7):ytad274. doi: 10.1093/ehjcr/ytad274. eCollection 2023 Jul.
7
Pseudoaneurysm Formation After "Preclose"-Assisted Impella Insertion in a Patient With Cardiogenic Shock.
J Med Cases. 2022 May;13(5):202-206. doi: 10.14740/jmc3841. Epub 2022 Apr 23.
9
How to Do It: Transbrachial Impella Placement to Facilitate Percutaneous Coronary Intervention.
Innovations (Phila). 2023 Jul-Aug;18(4):372-375. doi: 10.1177/15569845231190870. Epub 2023 Aug 8.

本文引用的文献

3
Impella support as a bridge to heart surgery in patients with cardiogenic shock.
Interact Cardiovasc Thorac Surg. 2022 Jul 9;35(2). doi: 10.1093/icvts/ivac088.
4
Coronary artery bypass grafting under sole Impella 5.0 support for patients with severely depressed left ventricular function.
J Artif Organs. 2022 Jun;25(2):158-162. doi: 10.1007/s10047-021-01285-1. Epub 2021 Jun 24.
5
Use of Impella CP Device in Off-Pump Coronary Artery Bypass Graft Surgery.
Int Heart J. 2021 Jan 30;62(1):175-177. doi: 10.1536/ihj.20-482. Epub 2021 Jan 16.
6
PCI Strategies in Patients with Acute Myocardial Infarction and Cardiogenic Shock.
N Engl J Med. 2017 Dec 21;377(25):2419-2432. doi: 10.1056/NEJMoa1710261. Epub 2017 Oct 30.
7
Percutaneous Mechanical Circulatory Support Versus Intra-Aortic Balloon Pump for Treating Cardiogenic Shock: Meta-Analysis.
J Am Coll Cardiol. 2017 Jan 24;69(3):358-360. doi: 10.1016/j.jacc.2016.10.026. Epub 2016 Oct 31.
8
Successful use of the Impella Recover LP 5.0 device for circulatory support during off-pump coronary artery bypass grafting.
Int J Surg Case Rep. 2014;5(11):803-5. doi: 10.1016/j.ijscr.2014.07.013. Epub 2014 Sep 22.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验