Departments of1Neurologic Surgery.
2Medical Scientist Training Program, Mayo Clinic, Rochester, Minnesota.
J Neurosurg. 2024 Mar 15;141(2):406-411. doi: 10.3171/2024.1.JNS232563. Print 2024 Aug 1.
Conventional frame-based stereotactic systems have circumferential base frames, often necessitating deep brain stimulation (DBS) surgery in two stages: intracranial electrode insertion followed by surgical re-preparation and pulse generator implantation. Some patients do not tolerate awake surgery, underscoring the need for a safe alternative for asleep DBS surgery. A frame-based stereotactic system with a skull-mounted "key" in lieu of a circumferential base frame received US FDA clearance. The authors describe the system's application for single-stage, asleep DBS surgery in 8 patients at their institution and review its workflow and technical considerations.
Eight patients underwent DBS lead insertion and IPG implantation in a single surgical preparation under general anesthesia using the system. Postoperative CT imaging confirmed lead placement.
Eight patients underwent implantation of 15 total leads targeting the ventral intermediate nucleus (4 patients), globus pallidus internus (GPi; 3 patients), and subthalamic nucleus (STN; 1 patient). Intraoperative microelectrode recording was conducted for GPi and STN targets. Postoperative CT imaging revealed a mean ± SD radial error of 1.24 ± 0.45 mm (n = 15 leads), without surgical complications.
The stereotactic system facilitated safe and effective asleep, single-stage DBS surgery, maintaining traditional lead accuracy standards.
传统的基于框架的立体定向系统具有环形底座框架,通常需要分两阶段进行深部脑刺激(DBS)手术:颅内电极插入,然后进行手术重新准备和脉冲发生器植入。一些患者无法耐受清醒手术,这凸显了需要一种安全的替代方法来进行睡眠 DBS 手术。一种带有颅骨安装“钥匙”的基于框架的立体定向系统已获得美国 FDA 批准。作者描述了该系统在其机构的 8 名患者中单阶段、睡眠 DBS 手术中的应用,并回顾了其工作流程和技术注意事项。
8 名患者在全身麻醉下使用该系统进行单次手术准备,进行 DBS 导联插入和 IPG 植入。术后 CT 成像确认了导联的位置。
8 名患者共植入了 15 根总导联,目标是腹侧中间核(4 名患者)、苍白球内(3 名患者)和丘脑底核(1 名患者)。对 GPi 和 STN 靶点进行了术中微电极记录。术后 CT 成像显示 15 根导联的平均±标准差径向误差为 1.24±0.45mm(n=15 根导联),无手术并发症。
该立体定向系统促进了安全有效的睡眠、单阶段 DBS 手术,保持了传统导联的准确性标准。