Department of Radiological Technology, Yamaguchi University Hospital, Minamikogushi, Yamaguchi, 755-8505, Japan.
Radiol Phys Technol. 2024 Jun;17(2):396-401. doi: 10.1007/s12194-024-00789-8. Epub 2024 Mar 22.
This study investigates the feasibility of estimating the radioactivity of radiopharmaceuticals using shielded syringes. The radioactivities of Tc-MDP, Tc-HMDP, Tc-ECD, Tc-MAG, and I-IMP were measured using a dose calibrator. Correlation coefficients and regression equations were obtained from the radioactivity in the shielded and unshielded syringes. Tc-MDP was also measured for residual radioactivity after the administration. The correlation coefficients of Tc-MDP, Tc-HMDP, Tc-ECD, Tc-MAG, and I-IMP were r = 0.9998, r = 0.9997, r = 0.9999, r = 0.9998, and r = 0.9888, respectively. The regression equations were y = 0.0364x + 0.0913, y = 0.0349x + 0.0273, y = 0.0343x - 0.0018, y = 0.0522x + 0.1215, and y = 0.0383x + 0.0058, respectively. The correlation coefficient for the residual radioactivity of Tc-MDP was r = 0.9887 and the regression equation was y = 0.1505x + 0.0853. The radioactivity of Tc- and I-labeled radiopharmaceuticals in shielded syringes was accurately measured. It was suggested that the measuring shielded syringes could provide an estimate of the actual radioactivity.
本研究探讨了使用屏蔽注射器估算放射性药物放射性的可行性。使用剂量计测量了 Tc-MDP、Tc-HMDP、Tc-ECD、Tc-MAG 和 I-IMP 的放射性。从屏蔽和未屏蔽注射器中的放射性获得了相关系数和回归方程。还测量了 Tc-MDP 给药后的放射性残留。Tc-MDP、Tc-HMDP、Tc-ECD、Tc-MAG 和 I-IMP 的相关系数分别为 r=0.9998、r=0.9997、r=0.9999、r=0.9998 和 r=0.9888。回归方程分别为 y=0.0364x+0.0913、y=0.0349x+0.0273、y=0.0343x-0.0018、y=0.0522x+0.1215 和 y=0.0383x+0.0058。Tc-MDP 放射性残留的相关系数为 r=0.9887,回归方程为 y=0.1505x+0.0853。准确测量了 Tc 和 I 标记放射性药物在屏蔽注射器中的放射性。建议测量屏蔽注射器可以提供实际放射性的估计。