Cardiovasc Revasc Med. 2024 Jul;64:70-75. doi: 10.1016/j.carrev.2024.02.015. Epub 2024 Mar 12.
This study evaluated the efficacy of a novel comprehensive shield designed to minimize radiation exposure (RE) to Physicians performing coronary and structural heart procedures.
The Protego™ radiation shielding system (Image Diagnostics Inc., Fitchburg, Ma) is designed to provide comprehensive protection from RE and has been State certified sufficient to allow operators to perform procedures without orthopedically burdensome lead aprons.
This single center two-group cohort study assessed the efficacy of this shield in a large number of cardiac procedures (coronary and structural), comparing operator RE compared to standard protection methods (personal lead apparel and "drop down" shield).
The Protego™ system reduced operator RE by 99 % compared to Standard Protection. RE was significantly lower at both "Head" level by thyroid median dose 0.0 (0.0, 0,0) vs 5.7 (2.9, 8.2) μSv (p < 0.001), as well as waist dose 0.0 (0.0, 0.0) vs 10.0 (5.0, 16.6) μSv (p < 0.001). "Zero" Total RE was documented by Raysafe™ in 64 % (n = 32) of TAVR cases and 73.2 % (n = 183) of the coronary cases utilizing Protego™. In contrast, standard protection did not achieve "Zero" exposure in a single case. These dramatic differences in RE were achieved despite higher fluoroscopy times in the Protego™ arm (11.9 ± 8.6 vs 14.3 ± 12.5 min, p = 0.015). Per case procedural exposure measured by Dose Area Product was higher in the Protego™ group compared to standard protection (115.4 ± 139.2 vs 74.9 ± 69.3, p < 0.001).
The Protego™ shield provides total body RE protection for operators performing both coronary and structural heart procedures. This shield allows procedural performance without the need for personal lead aprons and has potential to reduce catheterization laboratory occupational health hazards.
本研究评估了一种新型综合屏蔽装置在减少行冠状动脉和结构性心脏介入术的医生辐射暴露(RE)方面的效果。
Protego™ 辐射屏蔽系统(Image Diagnostics Inc.,马萨诸塞州菲特堡)旨在提供全面的保护,以避免 RE,且已通过国家认证,足以使操作人员在不佩戴笨重的铅围裙的情况下进行手术。
这项单中心的两组队列研究评估了这种屏蔽在大量心脏介入手术(冠状动脉和结构性)中的效果,将操作人员的 RE 与标准防护方法(个人铅衣和“下拉”屏蔽)进行了比较。
与标准防护相比,Protego™ 系统使操作人员的 RE 降低了 99%。在“头部”水平,甲状腺中位数剂量分别为 0.0(0.0,0,0)vs 5.7(2.9,8.2)μSv(p<0.001)和腰部剂量 0.0(0.0,0.0)vs 10.0(5.0,16.6)μSv(p<0.001)时,RE 显著降低。在使用 Protego™ 的情况下,Raysafe™ 记录到 64%(n=32)的 TAVR 病例和 73.2%(n=183)的冠状动脉病例实现了“零”总 RE。相比之下,标准防护在单个病例中均未达到“零”暴露。尽管在 Protego™ 组中透视时间更长(11.9±8.6 与 14.3±12.5 分钟,p=0.015),但仍取得了如此显著的 RE 差异。按例数计算的剂量面积乘积(DAP)测量的每个病例程序暴露量在 Protego™ 组中高于标准防护组(115.4±139.2 与 74.9±69.3,p<0.001)。
Protego™ 屏蔽装置为行冠状动脉和结构性心脏介入术的操作人员提供全身 RE 保护。该屏蔽装置允许在不使用个人铅围裙的情况下进行手术,并有可能降低导管室的职业健康危害。