Çaylak Osman
Department of Pharmacy Services, Vocational School of Health, Sivas Cumhuriyet University, Sivas, Turkiye.
Turk J Chem. 2023 Oct 29;48(1):21-35. doi: 10.55730/1300-0527.3636. eCollection 2024.
The chromium speciation procedure was optimized using magnetic polyaniline nanoparticles (FeO@PANI NPs) solid-phase extraction coupled with microsampling injection-flame atomic absorption spectrometry (MIS-FAAS). Chromium speciation was successfully achieved by FeO@PANI NPs at pH 8.0. The recoveries obtained for Cr(III) and Cr(VI) were above 95% and under 5%, respectively. Recoveries of over 95% for Cr(III) from 40.0 mL of the sample were obtained using 25 mg FeO@PANI NPs and 500 μL of 0.2% (w/v) thiourea (TU) solution prepared in 2 mol L HCl as eluent. Total chromium as Cr(III) was extracted quantitatively after reducing the Cr(VI) to Cr(III). The linear range, detection limit, preconcentration factor, and precision of the optimized method for Cr(III) in aqueous solution were 2.5-94.0 μg L, 0.335 μg L, 80, and 3.07%, respectively. The validation of the method was controlled using SPS-WW2 Batch 114 wastewater and BCR-715 industrial wastewater as standard reference materials (SRMs) for environmental water, and the obtained results were in close agreement with the certified values.
采用磁性聚苯胺纳米颗粒(FeO@PANI NPs)固相萃取结合微量进样注射-火焰原子吸收光谱法(MIS-FAAS)对铬形态分析方法进行了优化。通过FeO@PANI NPs在pH 8.0条件下成功实现了铬形态分析。Cr(III)和Cr(VI)的回收率分别高于95%和低于5%。使用25 mg FeO@PANI NPs和500 μL在2 mol L HCl中配制的0.2%(w/v)硫脲(TU)溶液作为洗脱剂,从40.0 mL样品中获得的Cr(III)回收率超过95%。将Cr(VI)还原为Cr(III)后,总铬以Cr(III)形式被定量萃取。该优化方法对水溶液中Cr(III)的线性范围、检测限、预富集因子和精密度分别为2.5 - 94.0 μg L、0.335 μg L、80和3.07%。使用SPS-WW2批次114废水和BCR-715工业废水作为环境水的标准参考物质(SRM)对该方法进行验证,所得结果与认定值高度吻合。