Biomechanical Characteristics of Glenosphere Orientation Based on Tilting Angle and Overhang Changes in Reverse Shoulder Arthroplasty.

机构信息

Department of Orthopaedic Surgery, Inje University Ilsan Paik Hospital, Goyang, Korea.

Department of Orthopaedic Surgery, National Health Insurance Service Ilsan Hospital, Goyang, Korea.

出版信息

Clin Orthop Surg. 2024 Apr;16(2):303-312. doi: 10.4055/cios23217. Epub 2024 Mar 15.

Abstract

BACKGROUND

Glenoid position and inclination are important factors in protecting against scapular notching, which is the most common complication that directly affects the longevity of reverse shoulder arthroplasty (RSA). This study aimed to investigate the biomechanical characteristics of glenosphere orientation, comparing neutral tilt, inferior overhang with an eccentric glenosphere at the same placement of baseplate, and inferior tilt after 10° inferior reaming in the lower part of the glenoid in RSA.

METHODS

Nine cadaveric shoulders were tested with 5 combinations of customized glenoid components: a centric glenosphere was combined with a standard baseplate (group A); an eccentric glenosphere to provide 4-mm inferior overhang than the centric glenosphere was combined with a standard baseplate (group B); a centric glenosphere was combined with a wedge-shaped baseplate tilted inferiorly by 10° with the same center of rotation (group C); an eccentric glenosphere was attached to a wedge-shaped baseplate (group D); and 10° inferior reaming was performed on the lower part of the glenoid to apply 10° inferior tilt, with a centric glenosphere secured to the standard baseplate for simulation of clinical tilt (group E). Impingement-free angles for adduction, abduction, forward flexion, external rotation, and internal rotation were measured. The capability of the deltoid moment arm for abduction and forward flexion, deltoid length, and geometric analysis for adduction engagement were evaluated.

RESULTS

Compared with neutral tilt, inferior tilt at the same position showed no significant difference in impingement-free angle, moment arm capability, and deltoid length. However, group D resulted in better biomechanical properties than a central position, regardless of inferior tilt. Group E demonstrated a greater range of adduction, internal and external rotation, and higher abduction and forward flexion capability with distalization, compared to corresponding parameters for inferior tilt with a customized wedge-shaped baseplate.

CONCLUSIONS

A 10° inferior tilt of the glenosphere, without changing the position of the baseplate, had no benefit in terms of the impingement-free angle and deltoid moment arm. However, an eccentric glenosphere had a significant advantage, regardless of inferior tilt. Inferior tilt through 10° inferior reaming showed better biomechanical results than neutral tilt due to the distalization effect.

摘要

背景

肩盂位置和倾斜度是预防肩峰切迹的重要因素,而肩峰切迹是影响反肩置换术(RSA)使用寿命的最常见并发症。本研究旨在探讨在 RSA 中,在同一肩胛盂假体位置下,比较中性倾斜、偏心假体下的下悬和偏心假体下的下倾,以及肩胛盂下部 10°下磨除后对肱骨头的生物力学影响。

方法

对 9 具尸体肩关节进行了 5 种定制肩胛盂假体的组合测试:中心型肱骨头假体联合标准基底部(A 组);偏心型肱骨头假体提供比中心型肱骨头假体 4mm 下悬的假体联合标准基底部(B 组);中心型肱骨头假体联合具有相同旋转中心的 10°下倾斜楔形基底部(C 组);偏心型肱骨头假体联合楔形基底部(D 组);在肩胛盂下部进行 10°下磨除,应用 10°下倾斜,中心型肱骨头假体固定于标准基底部以模拟临床倾斜(E 组)。测量内收、外展、前屈、外旋和内旋的无撞击角度。评估外展和前屈的三角肌力臂能力、三角肌长度和内收啮合的几何分析。

结果

与中性倾斜相比,同一位置的下倾斜在无撞击角度、力臂能力和三角肌长度方面没有显著差异。然而,无论下倾斜如何,D 组的生物力学性能均优于中心位置。与相应的定制楔形基底部下倾参数相比,E 组在外展、内旋和外旋以及更高的外展和前屈能力方面具有更大的范围,并且随着远移,具有更大的内收角度。

结论

不改变基底部位置的肱骨头 10°下倾斜在无撞击角度和三角肌力臂方面没有益处。然而,偏心型肱骨头假体具有显著优势,而与下倾斜无关。通过 10°下磨除肩胛盂下部进行下倾斜可获得更好的生物力学效果,这是由于远移效应。