Svanaes D B, Larheim T A, Backe S
Scand J Dent Res. 1985 Feb;93(1):61-7. doi: 10.1111/j.1600-0722.1985.tb01310.x.
A substantial dose reduction may be obtained for patients in need of frequent retakes with rotational panoramic radiography by reducing the width and height of the film area irradiated. An electronic timer was connected to the circuit of an Orthopantomograph 5 (Siemens) whereby exposure times could easily be selected in the range 5-15 s. Vertical extracollimation was provided. With the field size reduced to encompass the developing dentition only, about 60% reduction in integral absorbed dose was obtained using standard (Siemens special) screens. With rare earth (Titan HS) screens a reduction of 85% was obtained in integral absorbed dose and absorbed doses to the thyroid gland and eye lens were reduced by 70-79% and 80-89%, respectively. The doses thus obtained were equal to or less than those reported with a single dental film in speed Group D.
对于需要频繁重新拍摄旋转全景X线片的患者,通过减小照射的胶片区域的宽度和高度,可以大幅降低剂量。将一个电子定时器连接到Orthopantomograph 5(西门子)的电路上,从而可以轻松地在5 - 15秒范围内选择曝光时间。提供了垂直准直。当视野大小减小到仅覆盖正在发育的牙列时,使用标准(西门子专用)增感屏可使积分吸收剂量降低约60%。使用稀土(Titan HS)增感屏时,积分吸收剂量降低了85%,甲状腺和晶状体的吸收剂量分别降低了70 - 79%和80 - 89%。由此获得的剂量等于或低于速度组D中一张牙科胶片所报告的剂量。