Musculoskelet Sci Pract. 2024 Jun;71:102950. doi: 10.1016/j.msksp.2024.102950. Epub 2024 Mar 29.
The CROM instrument is widely used clinically and in research to measure neck range of motion. However, its measurement proprieties during the assessment of protraction and retraction movements were not examined so far.
To analyse the intra- and inter-rater reliability, the concurrent validity of the CROM for measuring head posture, retraction and protraction in healthy subjects.
Thirty-three asymptomatic subjects were recruited and assigned in a random order to one of two raters. After a 10-min break, they were examined by a second rater (Assessment 1). After a 30-min break, both raters repeated the examination (Assessment 2). The examination consisted of measuring the head posture, maximum head protraction and maximum retraction. Each movement was repeated 3 times and measured simultaneously with the CROM and with a 3D capture system laboratory.
The intra-rater reliability of the CROM was excellent for both raters for head posture and all head movements (ICC>0.9, 95% CI: 0.82-0.99, p < 0.01). The inter-rater reliability was excellent for head posture (ICC>0.95, 95% CI: 0.92-0.98, p < 0.01) and good-to-excellent for all movements at both time-points (ICC = 0.73-0.98, 95%CI: 0.45-0.99, p < 0.01). The validity analysis showed moderate-to-strong correlation between instruments for the head posture and head movements [(r) = -0.47 to -0.78), 95% CI: 0.99 to -0.24, p < 0.01].
The CROM instrument has good-to-excellent reliability and adequate validity for measuring cervical position and displacement in the sagittal plane.
CROM 仪器广泛应用于临床和研究中,用于测量颈部活动范围。然而,迄今为止,尚未对其在评估前伸和回缩运动时的测量特性进行检查。
分析 CROM 在测量健康受试者头部姿势、回缩和前伸时的内部和外部可靠性以及同时效度。
招募了 33 名无症状受试者,并按照随机顺序将他们分配给两名评估者之一。休息 10 分钟后,由第二名评估者进行检查(评估 1)。休息 30 分钟后,两名评估者重复检查(评估 2)。检查包括测量头部姿势、最大头部前伸和最大回缩。每个动作重复 3 次,同时使用 CROM 和 3D 捕获系统实验室进行测量。
两名评估者对头部姿势和所有头部运动的 CROM 内部可靠性均为优秀(ICC>0.9,95%CI:0.82-0.99,p<0.01)。评估者间可靠性对头部姿势为优秀(ICC>0.95,95%CI:0.92-0.98,p<0.01),对所有运动在两个时间点均为良好至优秀(ICC=0.73-0.98,95%CI:0.45-0.99,p<0.01)。有效性分析显示,头部姿势和头部运动的仪器之间存在中度至强相关性[(r)=-0.47 至-0.78),95%CI:0.99 至-0.24,p<0.01]。
CROM 仪器在测量矢状面上颈椎位置和位移方面具有良好至优秀的可靠性和足够的有效性。