Tosaka Kengo, Ninomiya Ryo, Fusazaki Tetsuya, Morino Yoshihiro
Division of Cardiology, Department of Internal Medicine, Iwate Medical University, Iwate, Japan.
J Cardiol Cases. 2024 Feb 16;29(4):190-192. doi: 10.1016/j.jccase.2023.12.008. eCollection 2024 Apr.
An 89-year-old man with symptomatic severe aortic stenosis underwent transcatheter aortic valve implantation due to old age and a history of coronary artery bypass grafting. Computed tomography showed a tricuspid aortic valve and severe calcification at the aortic valve annulus, with a perimeter of 88.7 mm. The 34-mm Evolut PRO+ (Medtronic Inc., Minneapolis, MN, USA) was selected. After balloon aortic valvuloplasty, deployment of the Evolut PRO+ was attempted, but significant expansion failure was observed. Upon retraction and removal of the Evolut PRO+ from the body, frame deformation was observed. A new Evolut PRO+ was tried again, but a similar finding was noted as a magatama-like infolding on transesophageal echocardiography. Fortunately, the patient's hemodynamics were relatively stable. Post-dilation was performed using a 25 mm -MED II (NuMED, Inc., Montreal, Canada) for reshaping.
In self-expanding transcatheter aortic valves (TAVs), bending of the TAV frame is widely known as one of the key problems. However, this is rare and infrequently encountered. In this case, TAV frame infolding occurred repeatedly, and the morphology of the infolding was evaluated in vitro and in vivo. Furthermore, we report that some TAVs can be reshaped by post-dilation.
一名89岁有症状的严重主动脉瓣狭窄男性患者,因年龄较大且有冠状动脉旁路移植术病史,接受了经导管主动脉瓣植入术。计算机断层扫描显示为三尖瓣主动脉瓣,主动脉瓣环严重钙化,周长为88.7毫米。选用了34毫米的Evolut PRO+(美敦力公司,美国明尼苏达州明尼阿波利斯)。在球囊主动脉瓣成形术后,尝试植入Evolut PRO+,但观察到明显的扩张失败。当将Evolut PRO+从体内撤回并取出时,观察到框架变形。再次尝试使用新的Evolut PRO+,但经食管超声心动图显示有类似勾玉样的折叠现象。幸运的是,患者的血流动力学相对稳定。使用25毫米的MED II(NuMED公司,加拿大蒙特利尔)进行后扩张以重塑瓣膜。
在自膨胀式经导管主动脉瓣(TAV)中,TAV框架弯曲是一个广为人知的关键问题。然而,这种情况很少见且不常遇到。在本病例中,TAV框架反复出现折叠,并在体外和体内对折叠形态进行了评估。此外,我们报告了一些TAV可通过后扩张进行重塑。