经导管主动脉瓣植入术中Evolut PRO+瓣膜折叠及球囊扩张后补救治疗:一例病例报告

Infolding of Evolut PRO+ during transcatheter aortic valve implantation and bailout by post-balloon dilation: A case report.

作者信息

Tosaka Kengo, Ninomiya Ryo, Fusazaki Tetsuya, Morino Yoshihiro

机构信息

Division of Cardiology, Department of Internal Medicine, Iwate Medical University, Iwate, Japan.

出版信息

J Cardiol Cases. 2024 Feb 16;29(4):190-192. doi: 10.1016/j.jccase.2023.12.008. eCollection 2024 Apr.

Abstract

UNLABELLED

An 89-year-old man with symptomatic severe aortic stenosis underwent transcatheter aortic valve implantation due to old age and a history of coronary artery bypass grafting. Computed tomography showed a tricuspid aortic valve and severe calcification at the aortic valve annulus, with a perimeter of 88.7 mm. The 34-mm Evolut PRO+ (Medtronic Inc., Minneapolis, MN, USA) was selected. After balloon aortic valvuloplasty, deployment of the Evolut PRO+ was attempted, but significant expansion failure was observed. Upon retraction and removal of the Evolut PRO+ from the body, frame deformation was observed. A new Evolut PRO+ was tried again, but a similar finding was noted as a magatama-like infolding on transesophageal echocardiography. Fortunately, the patient's hemodynamics were relatively stable. Post-dilation was performed using a 25 mm -MED II (NuMED, Inc., Montreal, Canada) for reshaping.

LEARNING OBJECTIVE

In self-expanding transcatheter aortic valves (TAVs), bending of the TAV frame is widely known as one of the key problems. However, this is rare and infrequently encountered. In this case, TAV frame infolding occurred repeatedly, and the morphology of the infolding was evaluated in vitro and in vivo. Furthermore, we report that some TAVs can be reshaped by post-dilation.

摘要

未标注

一名89岁有症状的严重主动脉瓣狭窄男性患者,因年龄较大且有冠状动脉旁路移植术病史,接受了经导管主动脉瓣植入术。计算机断层扫描显示为三尖瓣主动脉瓣,主动脉瓣环严重钙化,周长为88.7毫米。选用了34毫米的Evolut PRO+(美敦力公司,美国明尼苏达州明尼阿波利斯)。在球囊主动脉瓣成形术后,尝试植入Evolut PRO+,但观察到明显的扩张失败。当将Evolut PRO+从体内撤回并取出时,观察到框架变形。再次尝试使用新的Evolut PRO+,但经食管超声心动图显示有类似勾玉样的折叠现象。幸运的是,患者的血流动力学相对稳定。使用25毫米的MED II(NuMED公司,加拿大蒙特利尔)进行后扩张以重塑瓣膜。

学习目标

在自膨胀式经导管主动脉瓣(TAV)中,TAV框架弯曲是一个广为人知的关键问题。然而,这种情况很少见且不常遇到。在本病例中,TAV框架反复出现折叠,并在体外和体内对折叠形态进行了评估。此外,我们报告了一些TAV可通过后扩张进行重塑。

相似文献

1
Infolding of Evolut PRO+ during transcatheter aortic valve implantation and bailout by post-balloon dilation: A case report.
J Cardiol Cases. 2024 Feb 16;29(4):190-192. doi: 10.1016/j.jccase.2023.12.008. eCollection 2024 Apr.
2
Self-Expandable Transcatheter Aortic Valve Frame Infolding: An Increasingly Recognized Complication.
Eur J Case Rep Intern Med. 2020 Dec 14;7(12):002100. doi: 10.12890/2020_002100. eCollection 2020.
3
A word of caution using self-expanding transcatheter aortic valve-frame infolding.
Catheter Cardiovasc Interv. 2019 Feb 15;93(3):555-558. doi: 10.1002/ccd.27870. Epub 2018 Oct 30.
5
Balloon Versus Self-Expandable Valve for the Treatment of Bicuspid Aortic Valve Stenosis: Insights From the BEAT International Collaborative Registrys.
Circ Cardiovasc Interv. 2020 Jul;13(7):e008714. doi: 10.1161/CIRCINTERVENTIONS.119.008714. Epub 2020 Jul 10.
6
Self-expanding transcatheter aortic valve infolding: Current evidence, diagnosis, and management.
Catheter Cardiovasc Interv. 2021 Aug 1;98(2):E299-E305. doi: 10.1002/ccd.29432. Epub 2020 Dec 14.
7
Three Generations of Self-Expanding Transcatheter Aortic Valves: A Report From the STS/ACC TVT Registry.
JACC Cardiovasc Interv. 2020 Jan 27;13(2):170-179. doi: 10.1016/j.jcin.2019.08.035.
8
Outcomes of three different new generation transcatheter aortic valve prostheses.
Catheter Cardiovasc Interv. 2020 Feb 15;95(3):398-407. doi: 10.1002/ccd.28524. Epub 2019 Oct 14.

本文引用的文献

3
Self-expanding transcatheter aortic valve infolding: Current evidence, diagnosis, and management.
Catheter Cardiovasc Interv. 2021 Aug 1;98(2):E299-E305. doi: 10.1002/ccd.29432. Epub 2020 Dec 14.
5
Stent Frame Infolding of a 34-mm Evolut-R in a Patient With a Mechanical Mitral Valve.
JACC Cardiovasc Interv. 2020 Feb 10;13(3):e25-e27. doi: 10.1016/j.jcin.2019.10.009. Epub 2020 Jan 15.
6
A word of caution using self-expanding transcatheter aortic valve-frame infolding.
Catheter Cardiovasc Interv. 2019 Feb 15;93(3):555-558. doi: 10.1002/ccd.27870. Epub 2018 Oct 30.
7
Balloon Valvuloplasty for Evolut R Infolding: Useful Transesophageal Echocardiographic Monitoring for Diagnosis and Efficacy.
JACC Cardiovasc Interv. 2018 Sep 10;11(17):e135-e136. doi: 10.1016/j.jcin.2018.06.009. Epub 2018 Aug 15.
8
The "PAC-MAN" sign: an insidious complication after transcatheter aortic valve deployment.
Int J Cardiol. 2013 Sep 10;167(6):e192-4. doi: 10.1016/j.ijcard.2013.04.021. Epub 2013 Apr 28.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验