Higuchi Takashi, Taninaka Atsushi, Yoshimizu Rikuto, Hayashi Katsuhiro, Miwa Shinji, Yamamoto Norio, Tsuchiya Hiroyuki, Demura Satoru
Department of Orthopaedic Surgery, Kanazawa Red Cross Hospital, 2-251, Minma, Kanazawa 921-8162, Ishikawa, Japan.
Department of Orthopaedic Surgery, Graduate School of Medical Sciences, Kanazawa University, 13-1, Takaramachi, Kanazawa 921-8641, Ishikawa, Japan.
J Clin Med. 2024 Apr 19;13(8):2393. doi: 10.3390/jcm13082393.
: Recently, the T2 alpha nailing system (Stryker, Inc.), which has advanced locking screws that can attach a screw to a rod, has been used worldwide and is expected to improve fracture fixation. We analyzed two cases of supracondylar femoral fractures in older adult patients, in which intraoperative fractures occurred during the insertion of advanced locking screws of the T2 alpha femur retrograde intramedullary nail. : A 93-year-old and an 82-year-old woman each underwent T2 alpha femur retrograde nail fixation for supracondylar femur fractures at separate hospitals, and advanced locking screws were used as the proximal transverse locking screws. In both patients, a fracture line was observed at the proximal screw postoperatively, and the fractures were refixed with distal cable wiring and/or femoral distal plates. The patients were subsequently discharged from the same facility with no remarkable pain. : When inserting advanced locking screws, it is necessary to enlarge the screw hole in the near-bone cortex with a counterbore drill, which might add torque to the bone cortex that could result in fractures. If the sleeve is distant from the bone, the counterbore drill will not reach the bone, the screw hole will not expand, and the insertion of advanced locking screws will apply a strong torque to the bone cortex and may result in fracture. Moreover, it is important to confirm that the counterbore drill is securely inserted under fluoroscopy and to carefully enlarge the bony foramen manually to prevent fractures during screw insertion.
最近,T2阿尔法钉系统(史赛克公司)在全球范围内得到应用,该系统具有先进的锁定螺钉,可将螺钉固定在杆上,有望改善骨折固定效果。我们分析了两例老年患者的股骨髁上骨折病例,这两例患者在插入T2阿尔法股骨逆行髓内钉的先进锁定螺钉过程中发生了术中骨折。
一名93岁和一名82岁的女性分别在不同医院接受了T2阿尔法股骨逆行髓内钉固定治疗股骨髁上骨折,近端横向锁定螺钉采用先进的锁定螺钉。两名患者术后均在近端螺钉处观察到骨折线,骨折均采用远端缆线固定和/或股骨远端钢板重新固定。患者随后从同一医疗机构出院,无明显疼痛。
插入先进锁定螺钉时,需要用沉头钻扩大近骨皮质的螺钉孔,这可能会给骨皮质增加扭矩,从而导致骨折。如果套筒远离骨头,沉头钻将无法到达骨头,螺钉孔不会扩大,插入先进锁定螺钉会对骨皮质施加强大扭矩,可能导致骨折。此外,重要的是要在透视下确认沉头钻已牢固插入,并小心地手动扩大骨孔,以防止螺钉插入过程中发生骨折。