Bergeron Denis E, Cessna Jeffrey T, Broder Brittany A, Pibida Leticia, Fitzgerald Ryan P, DiGiorgio Morgan, Napoli Elisa, Zimmerman Brian E
Physical Measurement Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.
Physical Measurement Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.
Appl Radiat Isot. 2024 Jul;209:111326. doi: 10.1016/j.apradiso.2024.111326. Epub 2024 Apr 20.
Thorium-227 was separated from its progeny and standardized for activity by the triple-to-double coincidence ratio (TDCR) method of liquid scintillation counting. Confirmatory liquid scintillation-based measurements were made using efficiency tracing with H and live-timed anticoincidence counting (LTAC). The separation time and the efficiency of the separation were confirmed by gamma-ray spectrometry. Calibrations for reentrant pressurized ionization chambers, including commercial radionuclide calibrators, and a well-type NaI(Tl) detector are discussed.