测斜仪和应用程序比量角器更能准确测量前交叉韧带患者的膝关节伸展活动范围:三种设备的可靠性和最小可检测变化。

Inclinometers and Apps Are Better than Goniometers, Measuring Knee Extension Range of Motion in Anterior Cruciate Ligament Patients: Reliability and Minimal Detectable Change for the Three Devices.

机构信息

Rehabilitation Department, Aspetar, Orthopaedic and Sports Medicine Hospital, Doha, Qatar.

Department of Sports Medicine, Oslo Sports Trauma Research Center, Norwegian School of Sport Sciences, Oslo, Norway.

出版信息

J Knee Surg. 2024 Oct;37(12):821-827. doi: 10.1055/a-2321-0516. Epub 2024 May 7.

Abstract

Knee extension range of motion (ROM) measurement is important in patients with anterior cruciate ligament (ACL) injury. The main objective is to evaluate the reliability and the minimal detectable change (MDC) of three methods of measuring knee extension ROM in ACL patients. The three common goniometric devices were a universal goniometer, an inclinometer, and a smartphone app. During a single-visit, knee extension ROM was measured in both knees of 92 ACL-injured or -reconstructed patients by two testers blinded to the other's measures. Intrarater, interrater, and test-retest intraclass correlation coefficients (ICC) were calculated. Intrarater ICC was excellent for the three devices ranging from 0.92 to 0.94, with the inclinometer yielding the best results (ICC = 0.94 [95% confidence interval, CI: 0.91-0.96]). Interrater ICC, however, varied from 0.36 to 0.80. The inclinometer and the smartphone app yielded similar results 0.80 (95% CI: 0.71-0.86) and 0.79 (95% CI: 0.70-0.86), respectively, whereas the universal goniometer was 0.36 (95% CI: 0.17-0.53). Test-retest ICC for the inclinometer was 0.89 (95% CI: 0.84-0.93), 0.86 (95% CI: 0.79-0.91) for the app, and 0.83 (95% CI:0.74-0.89) for the goniometer. The intrarater, interrater, and test-retest MDC values ranged from 2.0 to 3.5, 3.7 to 10.4, and 2.6 to 5.4 degrees, respectively. The goniometer was the least reliable. The inclinometer is the recommended device due to its highest ICC scores among the three devices and ease of use.

摘要

膝关节伸展活动度(ROM)测量在 ACL 损伤患者中非常重要。主要目的是评估三种测量 ACL 患者膝关节伸展 ROM 方法的可靠性和最小可检测变化(MDC)。三种常见的测角器分别为通用测角器、测斜仪和智能手机应用程序。在单次就诊时,由两名测试员对 92 名 ACL 损伤或重建的患者的双膝进行测量,两名测试员均不知道对方的测量结果。计算了内部分离度、间部分离度和测试-再测试的组内相关系数(ICC)。三种设备的内部分离度 ICC 均为优秀,范围为 0.92 至 0.94,测斜仪的结果最佳(ICC=0.94 [95%置信区间,CI:0.91-0.96])。间部分离度 ICC 范围从 0.36 到 0.80。测斜仪和智能手机应用程序的结果相似,分别为 0.80(95% CI:0.71-0.86)和 0.79(95% CI:0.70-0.86),而通用测角器为 0.36(95% CI:0.17-0.53)。测斜仪的测试-再测试 ICC 为 0.89(95% CI:0.84-0.93),应用程序的 ICC 为 0.86(95% CI:0.79-0.91),测角器的 ICC 为 0.83(95% CI:0.74-0.89)。内部分离度、间部分离度和测试-再测试的 MDC 值范围分别为 2.0 至 3.5、3.7 至 10.4 和 2.6 至 5.4 度。测角器的可靠性最低。测斜仪是推荐的设备,因为它在三种设备中的 ICC 评分最高,且易于使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验