Suppr超能文献

使用单光子发射计算机断层扫描测定左心室质量。

Determination of left ventricular mass using single-photon emission computed tomography.

作者信息

Wolfe C L, Jansen D E, Corbett J R, Lipscomb K, Gabliani G, Filipchuk N, Redish G, Lewis S E, Willerson J T

出版信息

Am J Cardiol. 1985 Nov 1;56(12):761-4. doi: 10.1016/0002-9149(85)91131-2.

Abstract

To test the hypothesis that single-photon emission computed tomography (SPECT) could actually determine left ventricular LV mass in humans, SPECT measurements of LV mass were compared with LV mass determined by cineangiography in 12 patients with normal coronary arteries and LV function. Repeat SPECT determinations of LV mass were carried out in 5 patients. Projection images of the left ventricle were acquired after intravenous injection of thallium-201 (TI-201) using a rotating gamma camera. Transverse sections were reconstructed by filtered backprojection. The boundary of LV uptake of TI-201 in each transverse section was defined using a 3-dimensional threshold detector. Scintigraphic LV mass (total number of voxels demonstrating LV TI-201 uptake X voxel volume X specific gravity of myocardium) was compared with angiographic LV mass. There was good correlation between LV mass determined by SPECT and that determined by cineangiography. Mean angiographic LV mass was 208 +/- 45 g (+/- standard deviation). Mean SPECT LV mass was 204 +/- 42 g. Linear regression analysis revealed the following relation: SPECT LV mass = 0.76 X angiographic LV mass + 46.1 (r = 0.82, root-mean-square deviation from regression = 24.7). The SPECT values of LV mass varied an average of 10.4 +/- 4.6% (+/- standard deviation) in the 5 patients in whom 2 determinations were made. Thus, SPECT of TI-201 can accurately measure LV mass in humans.

摘要

为了验证单光子发射计算机断层扫描(SPECT)能否实际测定人体左心室(LV)质量这一假设,在12例冠状动脉和左心室功能正常的患者中,将SPECT测定的左心室质量与电影血管造影测定的左心室质量进行了比较。对5例患者进行了左心室质量的重复SPECT测定。静脉注射铊-201(TI-201)后,使用旋转γ相机采集左心室的投影图像。通过滤波反投影重建横断面图像。使用三维阈值探测器定义每个横断面中TI-201左心室摄取的边界。将闪烁扫描法测定的左心室质量(显示左心室TI-201摄取的体素总数×体素体积×心肌比重)与血管造影测定的左心室质量进行比较。SPECT测定的左心室质量与电影血管造影测定的左心室质量之间存在良好的相关性。平均血管造影左心室质量为208±45g(±标准差)。平均SPECT左心室质量为204±42g。线性回归分析显示以下关系:SPECT左心室质量 = 0.76×血管造影左心室质量 + 46.1(r = 0.82,回归的均方根偏差 = 24.7)。在进行了两次测定的5例患者中,左心室质量的SPECT值平均变化10.4±4.6%(±标准差)。因此,TI-201的SPECT能够准确测量人体的左心室质量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验