成人 99mTc-DTPA 动态 SPECT/CT 肾图:可行性和诊断获益。

99m Tc-DTPA dynamic SPECT/CT renogram in adults: feasibility and diagnostic benefit.

机构信息

Department of Nuclear Medicine, University Hospital of Patras, University of Patras, Medical School, Patras, Greece and .

Department of Physics, University of Bialystok, Bialystok, Poland, .

出版信息

Nucl Med Commun. 2024 Aug 1;45(8):673-682. doi: 10.1097/MNM.0000000000001865. Epub 2024 May 23.

Abstract

OBJECTIVE

The objective of this study is to investigate the feasibility and potential advantages of 99m Tc-DTPA dynamic single photon emission computed tomography/computed tomography (SPECT/CT) renogram in adults.

METHODS

Fifty-five patients aged 19-80 years (mean 56.3) were enrolled. The imaging protocol included: day 1: 99m Tc-DTPA planar renogram, followed by planar 99m Tc-DMSA scan. Day 3: attenuation-corrected dynamic 99m Tc-DTPA SPECT renogram [DSPECT(AC)] and Cr-51 ethylenediamine tetraacetic acid (EDTA) glomerular filtration rate (GFR) calculation. DSPECT(AC) included an initial CT scan followed by 12 consecutive SPECT sessions acquired via continuous-mode acquisition for a total of 24 min. Fast SPECT sequences (1-2 s/projection, 60 projections, every 6°) were obtained for the first 8 min, followed by slower acquisitions (3-4 s/projection) during the rest of the study. Renal activity was measured in the total kidney volume by regions of interest drawn on consecutive transaxial slices of the third SPECT, which were then copied on the whole 12-SPECT series. Corresponding time-activity curves were created. DSPECT(AC) parameters were compared with those of planar renogram. The reference method for split renal function was 99m Tc-DMSA (geometrical mean of anterior and posterior projection counts) and for GFR the Cr-51 EDTA 2-blood sample clearance method.

RESULTS

DSPECT(AC) images were of good quality. There was good correlation between renogram parameters (time to peak activity and NORA20) comparing the two techniques ( r  = 0.959 and 0.933, respectively). In 21 cases with >30% absolute difference between the two kidneys, spilt renal function calculation by DSPECT(AC) correlated perfectly ( r  = 0.968) with the reference method, whereas planar renogram was less accurate ( r  = 0.843). Anatomic information provided by nonenhanced CT offered an integrated structural-functional view valuable for final diagnosis. DSPECT(AC) early kidney uptake as a fraction of injected dose correlated better with reference GFR ( r  = 0.789) than the Gates' method ( r  = 0.642).

CONCLUSION

99m Tc-DTPA dynamic SPECT/CT renogram is feasible with conventional SPECT/CT systems. It allows accurate split renal function measurement, offers additional anatomical information and can be used for closer approximation of GFR compared with Gates' method.

摘要

目的

本研究旨在探讨 99mTc-DTPA 动态单光子发射计算机断层扫描/计算机断层扫描(SPECT/CT)肾图在成人中的可行性和潜在优势。

方法

纳入 55 名年龄在 19-80 岁(平均 56.3 岁)的患者。成像方案包括:第 1 天:99mTc-DTPA 平面肾图,随后进行 99mTc-DMSA 平面扫描。第 3 天:进行校正衰减的 99mTc-DTPA 动态 SPECT 肾图[DSPECT(AC)]和 Cr-51 乙二胺四乙酸(EDTA)肾小球滤过率(GFR)计算。DSPECT(AC)包括初始 CT 扫描,随后进行 12 次连续 SPECT 采集,通过连续模式采集共 24 分钟。前 8 分钟采集快速 SPECT 序列(1-2 秒/投影,60 个投影,每 6°采集 1 次),然后在研究的其余时间采集较慢的采集(3-4 秒/投影)。在第 3 次 SPECT 的连续轴位切片上用感兴趣区测量肾脏总活性,然后将这些切片复制到整个 12 次 SPECT 系列中。创建相应的时间-活性曲线。将 DSPECT(AC)参数与平面肾图进行比较。分肾功能的参考方法是 99mTc-DMSA(前位和后位计数的几何平均值),GFR 的参考方法是 Cr-51 EDTA 2-血样清除法。

结果

DSPECT(AC)图像质量良好。两种技术之间的肾图参数(达峰时间和 NORA20)具有良好的相关性(分别为 r = 0.959 和 0.933)。在 21 例两个肾脏之间的绝对差异大于 30%的情况下,DSPECT(AC)的分肾功能计算与参考方法完全相关(r = 0.968),而平面肾图的准确性较低(r = 0.843)。非增强 CT 提供的解剖学信息提供了有价值的综合结构-功能视图,有助于最终诊断。DSPECT(AC)早期肾脏摄取的注射剂量分数与参考 GFR 的相关性更好(r = 0.789),而与 Gates 方法的相关性较差(r = 0.642)。

结论

使用常规 SPECT/CT 系统,99mTc-DTPA 动态 SPECT/CT 肾图是可行的。它可以准确测量分肾功能,提供额外的解剖学信息,并可用于更接近 Gates 方法的 GFR 测量。