Wood M L, Henkelman R M
Ontario Cancer Institute, Toronto, Canada.
Magn Reson Med. 1985 Dec;2(6):517-26. doi: 10.1002/mrm.1910020602.
A method has been developed to reduce a magnetic resonance image artifact arising from the Fourier transformation of data truncated in the phase-encoding direction. The discontinuity is eliminated between the phase of the complex data sample associated with the maximum negative phase-encoding gradient and the phase of the maximum positive gradient sample. Images retain sharper edges than if low-pass filtering is employed. The method is demonstrated on clinical images, then an explanation of the method is presented.