Suppr超能文献

磁共振成像中的截断伪影。

Truncation artifacts in magnetic resonance imaging.

作者信息

Wood M L, Henkelman R M

机构信息

Ontario Cancer Institute, Toronto, Canada.

出版信息

Magn Reson Med. 1985 Dec;2(6):517-26. doi: 10.1002/mrm.1910020602.

Abstract

A method has been developed to reduce a magnetic resonance image artifact arising from the Fourier transformation of data truncated in the phase-encoding direction. The discontinuity is eliminated between the phase of the complex data sample associated with the maximum negative phase-encoding gradient and the phase of the maximum positive gradient sample. Images retain sharper edges than if low-pass filtering is employed. The method is demonstrated on clinical images, then an explanation of the method is presented.

摘要

已开发出一种方法,用于减少因在相位编码方向上截断的数据进行傅里叶变换而产生的磁共振图像伪影。消除了与最大负相位编码梯度相关的复数数据样本的相位与最大正梯度样本的相位之间的不连续性。与采用低通滤波相比,图像保留了更清晰的边缘。该方法在临床图像上得到了验证,然后对该方法进行了解释。